An Endorectal Balloon Reduces Intrafraction Prostate Motion During Radiotherapy

Overview
Specialties Oncology
Radiology
Date 2011 Nov 22
PMID 22099035
Citations 40
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To investigate the effect of endorectal balloons (ERBs) on intrafraction and interfraction prostate motion during radiotherapy.

Methods And Materials: Thirty patients were treated with intensity-modulated radiotherapy, to a total dose of 80 Gy in 40 fractions. In 15 patients, a daily-inserted air-filled ERB was applied. Prostate motion was tracked, in real-time, using an electromagnetic tracking system. Interfraction displacements, measured before each treatment, were quantified by calculating the systematic and random deviations of the center of mass of the implanted transponders. Intrafraction motion was analyzed in timeframes of 150 s, and displacements >1 mm, >3 mm, >5 mm, and >7 mm were determined in the anteroposterior, left-right, and superoinferior direction, and for the three-dimensional (3D) vector. Manual table corrections, made during treatment sessions, were retrospectively undone.

Results: A total of 576 and 567 tracks have been analyzed in the no-ERB group and ERB group, respectively. Interfraction variation was not significantly different between both groups. After 600 s, 95% and 98% of the treatments were completed in the respective groups. Significantly fewer table corrections were performed during treatment fractions with ERB: 88 vs. 207 (p = 0.02). Intrafraction motion was significantly reduced with ERB. During the first 150 s, only negligible deviations were observed, but after 150 s, intrafraction deviations increased with time. This resulted in cumulative percentages of 3D-vector deviations >1 mm, >3 mm, >5 mm, and >7 mm that were 57.7%, 7.0%, 0.7%, and 0.3% in the ERB-group vs. 70.2%, 18.1%, 4.6%, and 1.4% in the no-ERB group after 600 s. The largest reductions in the ERB group were observed in the AP direction. These data suggest that a 5 mm CTV-to-PTV margin is sufficient to correct for intrafraction prostate movements when using an ERB.

Conclusions: ERB significantly reduces intrafraction prostate motion, but not interfraction variation, and may in particular be beneficial for treatment sessions longer than 150 s.

Citing Articles

Planning CT Identifies Patients at Risk of High Prostate Intrafraction Motion.

Ballhausen H, Li M, Lombardo E, Landry G, Belka C Cancers (Basel). 2023; 15(16).

PMID: 37627131 PMC: 10452220. DOI: 10.3390/cancers15164103.


Transperineal ultrasound is a good alternative for intra-fraction motion monitoring for prostate stereotactic body radiotherapy.

Guo B, Stephans K, Godley A, Kolar M, Magnelli A, Tendulkar R J Appl Clin Med Phys. 2023; 24(10):e14021.

PMID: 37144947 PMC: 10562017. DOI: 10.1002/acm2.14021.


Prostate stereotactic body radiotherapy: quantifying intra-fraction motion and calculating margins using the new BIR geometric uncertainties in daily online IGRT recommendations.

McNeice J, Sanilkumar N, Alexander S, Talbot J, Tree A, McNair H Br J Radiol. 2023; 96(1146):20220852.

PMID: 37001054 PMC: 10230383. DOI: 10.1259/bjr.20220852.


Are simple verbal instructions sufficient to ensure that bladder volume does not deteriorate prostate position reproducibility during spot scanning proton therapy?.

Nishioka K, Gotoh K, Hashimoto T, Abe T, Osawa T, Matsumoto R BJR Open. 2022; 3(1):20210064.

PMID: 35707757 PMC: 9185850. DOI: 10.1259/bjro.20210064.


Urethra Sparing With Target Motion Mitigation in Dose-Escalated Extreme Hypofractionated Prostate Cancer Radiotherapy: 7-Year Results From a Phase II Study.

Greco C, Pares O, Pimentel N, Louro V, Nunes B, Kociolek J Front Oncol. 2022; 12:863655.

PMID: 35433469 PMC: 9012148. DOI: 10.3389/fonc.2022.863655.