Clinical Validation of the Acuros XB Photon Dose Calculation Algorithm, a Grid-based Boltzmann Equation Solver

Overview
Journal Acta Oncol
Specialty Oncology
Date 2011 Nov 9
PMID 22060134
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: A new algorithm that uses a grid-based technique to solve the linear Boltzmann transport equation (LBTE) has been developed to improve the accuracy and speed of external photon beam treatment planning calculations. The aim of this study was to test the accuracy of this algorithm in both heterogeneous and homogeneous media.

Material And Methods: Output factors, depth dose curves and profiles for symmetric fields were measured in water using diamond and ionization chamber detectors. Furthermore, asymmetric fields, fields collimated with the multi-leaf collimator, enhanced dynamic wedge fields as well as fields with different source-skin distances were measured. Various test plans were created on a CIRS thorax phantom including tissue-equivalent inserts and corresponding dose distributions within the phantom were measured with radiochromic films. The new grid-based LBTE solver, Acuros XB (Eclipse version 10.0, Varian Medical Systems, CA, USA) was used to calculate dose distributions for all field configurations and plans, for both 6 MV and 15 MV photons. Calculations were also performed with AAA, a standard convolution algorithm.

Results: Compared to measurements, the output factors were within 1% for Acuros XB. For the depth doses, the average deviations were within 1% in dose and 1 mm in distance to agreement (DTA). For the profiles, the deviations were within 2%/1 mm except near the penumbra. Similar results were obtained for the other field configurations. Good agreement with AAA was also found. For the plans calculated on the CIRS phantom, the number of points meeting the gamma criterion of 3% in dose and 3 mm in DTA was higher with Acuros XB (98% for 6 MV; 100% for 15 MV) than with AAA (94% for 6 MV; 96% for 15 MV).

Conclusion: Dose calculations with the Acuros XB algorithm in homogeneous media are in good agreement with both measurements and the AAA algorithm. In heterogeneous media, the Acuros XB algorithm is superior to AAA in both lung and bony material.

Citing Articles

Dosimetric Systems in Pre-Treatment QA for Stereotactic Treatments: Correlation Agreements and Target Volume Dependency.

Muthu S, Mudhana G Asian Pac J Cancer Prev. 2024; 25(4):1425-1432.

PMID: 38680004 PMC: 11162722. DOI: 10.31557/APJCP.2024.25.4.1425.


Are offline ART decisions for NSCLC impacted by the type of dose calculation algorithm?.

Callens D, Aerts K, Berkovic P, Vandewinckele L, Lambrecht M, Crijns W Tech Innov Patient Support Radiat Oncol. 2024; 29:100236.

PMID: 38313556 PMC: 10835600. DOI: 10.1016/j.tipsro.2024.100236.


Extending deterministic transport capabilities for very-high and ultra-high energy electron beams.

Naceur A, Bienvenue C, Romano P, Chilian C, Carrier J Sci Rep. 2024; 14(1):2796.

PMID: 38307920 PMC: 11226718. DOI: 10.1038/s41598-023-51143-8.


Feasibility of a multigroup Boltzmann-Fokker-Planck solution for electron beam dose calculations.

Naceur A, Hebert A, Romano P, Forget B, Chilian C, Carrier J Sci Rep. 2023; 13(1):1310.

PMID: 36693824 PMC: 9873679. DOI: 10.1038/s41598-023-27376-y.


Influence of optional measurement parameters in the Eclipse treatment planning system on the quality of the dosimetric model of the biomedical accelerator using the Acuros XB algorithm.

Gadek A, Wendykier J, Grzadziel A, Bekman B, Smolinska B, Bekman A Rep Pract Oncol Radiother. 2022; 27(2):241-249.

PMID: 36299384 PMC: 9591040. DOI: 10.5603/RPOR.a2022.0019.