» Articles » PMID: 21911539

A Comparison of Parallel and Diverging Screw Angles in the Stability of Locked Plate Constructs

Overview
Date 2011 Sep 14
PMID 21911539
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

We investigated the static and cyclical strength of parallel and angulated locking plate screws using rigid polyurethane foam (0.32 g/cm(3)) and bovine cancellous bone blocks. Custom-made stainless steel plates with two conically threaded screw holes with different angulations (parallel, 10° and 20° divergent) and 5 mm self-tapping locking screws underwent pull-out and cyclical pull and bending tests. The bovine cancellous blocks were only subjected to static pull-out testing. We also performed finite element analysis for the static pull-out test of the parallel and 20° configurations. In both the foam model and the bovine cancellous bone we found the significantly highest pull-out force for the parallel constructs. In the finite element analysis there was a 47% more damage in the 20° divergent constructs than in the parallel configuration. Under cyclical loading, the mean number of cycles to failure was significantly higher for the parallel group, followed by the 10° and 20° divergent configurations. In our laboratory setting we clearly showed the biomechanical disadvantage of a diverging locking screw angle under static and cyclical loading.

Citing Articles

Finite element method analysis of bone stress for variants of locking plate placement.

Palka M, Miszczyk P, Jurewicz M, Perz R Heliyon. 2024; 10(8):e26840.

PMID: 38660239 PMC: 11039970. DOI: 10.1016/j.heliyon.2024.e26840.


Preliminary study on a novel dedicated plate for iliac fractures in dogs.

Moi T, Minto B, Macedo A, Lucena D, Malta C, Dias L PLoS One. 2022; 17(8):e0269313.

PMID: 36018880 PMC: 9416997. DOI: 10.1371/journal.pone.0269313.


Comparing the locking screw direction of three locking plates for lateral clavicle fractures: a simulation study.

Abe S, Koizumi K, Murase T, Kuriyama K BMC Musculoskelet Disord. 2021; 22(1):812.

PMID: 34548051 PMC: 8456609. DOI: 10.1186/s12891-021-04697-5.


Indications and results of osteosynthesis for proximal humerus fragility fractures in elderly patients.

Murena L, Canton G, Ratti C, Hoxhaj B, Giraldi G, Surace M Orthop Rev (Pavia). 2020; 12(1):8559.

PMID: 32391138 PMC: 7206366. DOI: 10.4081/or.2020.8559.


Complex biomechanical properties of non-augmented and augmented pedicle screws in human vertebrae with reduced bone density.

Schulze M, Riesenbeck O, Vordemvenne T, Raschke M, Evers J, Hartensuer R BMC Musculoskelet Disord. 2020; 21(1):151.

PMID: 32143605 PMC: 7060638. DOI: 10.1186/s12891-020-3158-z.