» Articles » PMID: 21492791

Microfluidic Preparation of [18F]FE@SUPPY and [18F]FE@SUPPY:2--comparison with Conventional Radiosyntheses

Overview
Journal Nucl Med Biol
Publisher Elsevier
Date 2011 Apr 16
PMID 21492791
Citations 7
Authors
Affiliations
Soon will be listed here.
Abstract

Introduction: Recently, first applications of microfluidic principles for radiosyntheses of positron emission tomography compounds were presented, but direct comparisons with conventional methods were still missing. Therefore, our aims were (1) the set-up of a microfluidic procedure for the preparation of the recently developed adenosine A(3)-receptor tracers [(18)F]FE@SUPPY [5-(2-[(18)F]fluoroethyl)2,4-diethyl-3-(ethylsulfanylcarbonyl)-6-phenylpyridine-5-carboxylate] and [(18)F]FE@SUPPY:2 [5-ethyl-2,4-diethyl-3-((2-[(18)F]fluoroethyl)sulfanylcarbonyl)-6-phenylpyridine-5-carboxylate] and (2) the direct comparison of reaction conditions and radiochemical yields of the no-carrier-added nucleophilic substitution with [(18)F]fluoride between microfluidic and conventional methods.

Methods: For the determination of optimal reaction conditions within an Advion NanoTek synthesizer, 5-50 μl of precursor and dried [(18)F]fluoride solution were simultaneously pushed through the temperature-controlled reactor (26 °C-180 °C) with defined reactant bolus flow rates (10-50 μl/min). Radiochemical incorporation yields (RCIYs) and overall radiochemical yields for large-scale preparations were compared with data from conventional batch-mode syntheses.

Results: Optimal reaction parameters for the microfluidic set-up were determined as follows: 170 °C, 30-μl/min pump rate per reactant (reaction overall flow rate of 60 μl/min) and 5-mg/ml precursor concentration in the reaction mixture. Applying these optimized conditions, we observed a significant increase in RCIY from 88.2% to 94.1% (P < .0001, n ≥ 11) for [(18)F]FE@SUPPY and that from 42.5% to 95.5% (P<.0001, n ≥ 5) for [(18)F]FE@SUPPY:2 using microfluidic instead of conventional heating. Precursor consumption was decreased from 7.5 and 10 mg to 1 mg per large-scale synthesis for both title compounds, respectively.

Conclusion: The direct comparison of radiosyntheses data applying a conventional method and a microfluidic approach revealed a significant increase of RCIY using the microfluidic approach.

Citing Articles

Development and implementation of ISAR, a new synthesis platform for radiopharmaceutical production.

Frank C, Winter G, Rensei F, Samper V, Brooks A, Hockley B EJNMMI Radiopharm Chem. 2019; 4(1):24.

PMID: 31659546 PMC: 6751239. DOI: 10.1186/s41181-019-0077-0.


The Current Role of Microfluidics in Radiofluorination Chemistry.

Knapp K, Nickels M, Manning H Mol Imaging Biol. 2019; 22(3):463-475.

PMID: 31485889 DOI: 10.1007/s11307-019-01414-6.


Preparation and First Preclinical Evaluation of [(18)F]FE@SNAP: A Potential PET Tracer for the Melanin-Concentrating Hormone Receptor-1 (MCHR1).

Philippe C, Nics L, Zeilinger M, Schirmer E, Spreitzer H, Karanikas G Sci Pharm. 2013; 81(3):625-39.

PMID: 24106662 PMC: 3791928. DOI: 10.3797/scipharm.1306-02.


Microfluidics: a groundbreaking technology for PET tracer production?.

Rensch C, Jackson A, Lindner S, Salvamoser R, Samper V, Riese S Molecules. 2013; 18(7):7930-56.

PMID: 23884128 PMC: 6270045. DOI: 10.3390/molecules18077930.


Plug-and-play modules for flexible radiosynthesis.

Herman H, Flores G, Quinn K, Eddings M, Olma S, Moore M Appl Radiat Isot. 2013; 78:113-24.

PMID: 23702795 PMC: 4492277. DOI: 10.1016/j.apradiso.2013.04.023.