Response to Cardiac Markers in Human Serum Analyzed by Guided-mode Resonance Biosensor

Overview
Journal Anal Chem
Specialty Chemistry
Date 2010 Nov 6
PMID 21049960
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Cardiac markers in human serum with concentrations less than 0.1 ng/mL were analyzed by use of a guided-mode resonance (GMR) biosensor. Cardiac troponin I (cTnI), creatine kinase MB (CK-MB), and myoglobin (MYO) were monitored in the serum of both patients and healthy controls. Dose-response curves ranging from 0.05 to 10 ng/mL for cTnI, from 0.1 to 10 ng/mL for CK-MB, and from 0.03 to 1.7 μg/mL for MYO were obtained. The limits of detection (LOD) for cTnI, CK-MB, and MYO were less than 0.05, 0.1, and 35 ng/mL, respectively. Analysis time was 30 min, which is short enough to meet clinical requirements. Antibody immobilization and the hydrophilic properties of the guided-mode resonance filter (GMRF) surface were investigated by X-ray photoelectron spectroscopy (XPS) and by monitoring the peak wavelength shift and water contact angle (CA). Both assays used to evaluate the surface density of the immobilized antibodies, a sandwich enzyme-linked immunosorbent assay (ELISA) and a sandwich immunogold assay, showed that the antibodies were successfully immobilized and sufficiently aligned to detect the low concentration of biomarkers. Our results show that the GMR biosensor will be very useful in developing low-cost portable biosensors that can screen for cardiac diseases.

Citing Articles

Point-of-Care Testing for Multiple Cardiac Markers Based on a Snail-Shaped Microfluidic Chip.

Yin B, Wan X, Qian C, Sohan A, Wang S, Zhou T Front Chem. 2021; 9:741058.

PMID: 34671590 PMC: 8521045. DOI: 10.3389/fchem.2021.741058.


Developing an Optical Interferometric Detection Method based biosensor for detecting specific SARS-CoV-2 immunoglobulins in Serum and Saliva, and their corresponding ELISA correlation.

Murillo A, Tome-Amat J, Ramirez Y, Garrido-Arandia M, Valle L, Hernandez-Ramirez G Sens Actuators B Chem. 2021; 345:130394.

PMID: 34248283 PMC: 8253720. DOI: 10.1016/j.snb.2021.130394.


Axially-Anisotropic Hierarchical Grating 2D Guided-Mode Resonance Strain-Sensor.

Babu S, Lee J Sensors (Basel). 2019; 19(23).

PMID: 31795120 PMC: 6929076. DOI: 10.3390/s19235223.


Automated Chemical Sensing Unit Integration for Parallel Optical Interrogation.

Hernandez A, Dortu F, Veenstra T, Ciaurriz P, Casquel R, Cornago I Sensors (Basel). 2019; 19(4).

PMID: 30791592 PMC: 6412770. DOI: 10.3390/s19040878.


Guided-mode resonant narrowband terahertz filtering by periodic metallic stripe and patch arrays on cyclo-olefin substrates.

Ferraro A, Zografopoulos D, Caputo R, Beccherelli R Sci Rep. 2018; 8(1):17272.

PMID: 30467397 PMC: 6250693. DOI: 10.1038/s41598-018-35515-z.