» Articles » PMID: 20588529

Full-range Polarization-sensitive Swept-source Optical Coherence Tomography by Simultaneous Transversal and Spectral Modulation

Overview
Journal Opt Express
Date 2010 Jul 1
PMID 20588529
Citations 17
Authors
Affiliations
Soon will be listed here.
Abstract

Polarization-sensitive swept-source optical coherence tomography (PS-SS-OCT) is used to measure three-dimensional phase-retardation images of birefringent biological tissue in vivo. PS-SS-OCT with continuous source polarization modulation is used to multiplex the incident states of polarization in the signal frequency of each A-scan. Although it offers the advantage of measurement speed that is as high as that of standard SS-OCT, its disadvantage is low axial measurement range. To overcome this drawback, we employed the B-M-mode scan (BM-scan) method, which removes complex conjugate ambiguity by applying phase modulation along the transversal scanning direction. Since polarization modulation and BM-scan are applied in different scanning directions, these methods can be combined to make the optimum use of both full range and polarization-sensitive imaging. Phase fluctuations that cause measurement failure were numerically canceled before demodulating the B-scan oriented modulation. After removing complex conjugate artifacts, the axial measurement range was 5.35 mm, and the signal-to-conjugate ratio was 40.5 dB. We demonstrated retinal imaging using the PS-SS-OCT system with a frequency-swept laser at a center wavelength of 1064 nm and an axial resolution of 11.4 microm in tissue. Full-range polarization-sensitive retinal images showed characteristic birefringence of fibrous tissues such as retinal nerve fiber, sclera, and lamina cribrosa.

Citing Articles

Full-range space-division multiplexing optical coherence tomography angiography.

Huang Y, Jerwick J, Liu G, Zhou C Biomed Opt Express. 2020; 11(8):4817-4834.

PMID: 32923080 PMC: 7449723. DOI: 10.1364/BOE.400162.


Parallel detection of Jones-matrix elements in polarization-sensitive optical coherence tomography.

Yamanari M, Uematsu S, Ishihara K, Ikuno Y Biomed Opt Express. 2019; 10(5):2318-2336.

PMID: 31149375 PMC: 6524579. DOI: 10.1364/BOE.10.002318.


detection of UV-induced acute skin effects using optical coherence tomography.

Chen W, Chang Y, Shen S, Tzeng Y, Lee H, Yang C Biomed Opt Express. 2019; 9(9):4235-4245.

PMID: 30615717 PMC: 6157765. DOI: 10.1364/BOE.9.004235.


Pixel-wise segmentation of severely pathologic retinal pigment epithelium and choroidal stroma using multi-contrast Jones matrix optical coherence tomography.

Azuma S, Makita S, Miyazawa A, Ikuno Y, Miura M, Yasuno Y Biomed Opt Express. 2018; 9(7):2955-2973.

PMID: 29984078 PMC: 6033570. DOI: 10.1364/BOE.9.002955.


Estimation of Jones matrix, birefringence and entropy using Cloude-Pottier decomposition in polarization-sensitive optical coherence tomography.

Yamanari M, Tsuda S, Kokubun T, Shiga Y, Omodaka K, Aizawa N Biomed Opt Express. 2016; 7(9):3551-3573.

PMID: 27699120 PMC: 5030032. DOI: 10.1364/BOE.7.003551.