Abstract

Background And Objective: The main objective of the present study is to demonstrate the feasibility of utilizing a novel non-invasive radiofrequency (RF) device to induce lethal thermal damage to subcutaneous adipose tissue only by establishing a controlled electric field that heats up fat preferentially.

Study Design/materials And Methods: Adipocyte cells in six-well plates were subjected to hyperthermic conditions: 45, 50, 55, 60, and 65 degrees C during 1, 2, and 3 minutes. Cell viability was assessed 72 hours after exposure. Two groups of abdominoplasty patients were treated with the RF device during and days before their surgical procedure. Temperatures of cutaneous and subcutaneous tissues were measured during treatment (3 minutes) of the first group. The immediate tissue response to heating was assessed by acute histology. The delayed tissue response was assessed by histology analysis of the second group, 4, 9, 10, 17, and 24 days after treatment (22 minutes). A mathematical model was used to estimate treatment temperatures of the second group. The model uses patient-based diagnostic measurements as input and was validated with in vivo clinical temperature measurements.

Results: Cell viability dropped from 89% to 20% when temperature increased from 45 to 50 degrees C during 1 minute exposures. Three minutes at 45 degrees C resulted in 40% viability. In vivo, the temperature of adipose tissue at 7-12 mm depth from the surface increased to 50 degrees C while the temperature of cutaneous tissues was <30 degrees C during RF exposure. Acute and longitudinal histology evaluations show normal epidermal and dermal layers. Subcutaneous tissues were also normal acutely. Subcutaneous vascular alterations, starting at day 4, and fat necrosis, starting at day 9, were consistently observed within 4.5-19 mm depth from the skin surface. Subcutaneous tissue temperatures were estimated to be 43-45 degrees C for 15 minutes.

Conclusions: A controlled internal electric field perpendicular to the skin-fat interface is selective in heating up fat and, consequently, has the ability to induce lethal thermal damage to subcutaneous adipose tissues while sparing overlying and underlying tissues. In vitro adipocyte cells are heat sensitive to thermal exposures of 50 and 45 degrees C on the order of minutes, 1 and 3 minutes, respectively. In vivo, 15 minutes thermal exposures to 43-45 degrees C result in a delayed adipocyte cellular death response-in this study, 9 days. The novel RF device presented herein effectively delivers therapeutic thermal exposures to subcutaneous adipose tissues while protecting epidermal and dermal layers.

Citing Articles

Histopathological Analysis of Abdominal Skin and Fat After In Vivo Hyperthermia-Induced Lipolysis With a 1064 nm Diode Laser, Radiofrequency, and Electromagnetic Field: A Pilot Study.

Nadarajan V, Panton J, Vingan N, Walmsley C, Barillas J, Lazzarini A Aesthet Surg J Open Forum. 2025; 7:ojae127.

PMID: 39959417 PMC: 11826928. DOI: 10.1093/asjof/ojae127.


Adipocyte Apoptosis Following a Novel Method for Double Chin Reduction: A Pilot Human Histology Study.

Goldberg D J Cosmet Dermatol. 2024; 24(1):e16643.

PMID: 39536780 PMC: 11743306. DOI: 10.1111/jocd.16643.


The Impact of the Number and Duration of Treatments With a 1064 nm Diode Laser on Adipocyte Apoptosis: Implications for Noninvasive Fat Reduction Strategies.

Salameh F, Lazzarini A, Kenkel J, Akgul Y Aesthet Surg J Open Forum. 2024; 6:ojae082.

PMID: 39512807 PMC: 11541045. DOI: 10.1093/asjof/ojae082.


Novel Applicator Utilizing HIFES and Enhanced Synchronized Radiofrequency+ for Subcutaneous Fat Reduction: Porcine Model Study.

Robb C, Bernardy J, Jarosova R, Hodkovicova N Lasers Surg Med. 2024; 56(10):803-810.

PMID: 39462846 PMC: 11629285. DOI: 10.1002/lsm.23854.


Efficacy of 1060 nm Diode Laser for Non-Invasive Subcutaneous Fat Reduction in Mini-Pigs.

Lee Y, Hong N, Choi E, Choi S, Kim E, Park H Clin Cosmet Investig Dermatol. 2024; 17:2097-2110.

PMID: 39318477 PMC: 11421435. DOI: 10.2147/CCID.S471367.