» Articles » PMID: 20580286

Biomechanical Evaluation of a T-shaped Miniplate Fixation of a Modified Sagittal Split Ramus Osteotomy with Buccal Step, a New Technique for Mandibular Orthognathic Surgery

Overview
Publisher Elsevier
Date 2010 Jun 29
PMID 20580286
Citations 6
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: The purpose of this study was to biomechanically evaluate the stability of a T-shaped miniplate fixation of a modified sagittal split ramus osteotomy (MSSRO) with buccal step and to compare it with single or double-parallel straight miniplates fixing a standard sagittal split ramus osteotomy (SSRO).

Study Design: Eighteen Synbone mandibular replicas were used in the study and divided into 3 groups. Standard SSRO was applied in the first and second groups, and the third group was cut for MSSRO with buccal step. After 7 mm of advancement, fixation modalities for the 3 groups included a single straight miniplate, double-parallel straight miniplates, and a T-shape miniplate, respectively. Each model was secured in a jig and subjected to vertical load on the anterior teeth.

Results: The T miniplate group showed a significantly higher value for stability than the group with a single straight miniplate. There was no significant difference in stability between the T miniplate and the double-parallel straight miniplate groups.

Conclusion: For mandibular advancement surgery of 7 mm in a laboratory environment, a T-shaped miniplate used with MSSRO and buccal step as a combination significantly optimize the resistance and stability of the fixation compared with a standard SSRO fixed with a single straight miniplate.

Citing Articles

Biomechanical evaluation of different fixation techniques following modified sagittal split ramus osteotomy using buccal step: An study.

Warade P, Kshirsagar R, Mishra P, Prasad P, Sardeshmukh A Natl J Maxillofac Surg. 2023; 13(3):443-448.

PMID: 36683918 PMC: 9851348. DOI: 10.4103/njms.njms_321_21.


Comparison of the Stability of Mandibular Sagittal Osteotomy Fixation between Two Types of Titanium Miniplates: A Biomechanical Study in Sheep Mandibles.

Trento G, Pires F, Dos Santos F, da Costa D, Rebellato N, Kluppel L Craniomaxillofac Trauma Reconstr. 2018; 11(2):114-117.

PMID: 29892325 PMC: 5993659. DOI: 10.1055/s-0037-1603455.


Biomechanical Loading Evaluation of Unsintered Hydroxyapatite/poly-l-lactide Plate System in Bilateral Sagittal Split Ramus Osteotomy.

Sukegawa S, Kanno T, Manabe Y, Matsumoto K, Sukegawa-Takahashi Y, Masui M Materials (Basel). 2017; 10(7).

PMID: 28773126 PMC: 5551807. DOI: 10.3390/ma10070764.


Analysis of Mandibular Test Specimens Used to Assess a Bone Fixation System.

Baccarin L, Casarin R, Lopes-da-Silva J, Passeri L Craniomaxillofac Trauma Reconstr. 2015; 8(3):171-8.

PMID: 26269724 PMC: 4532570. DOI: 10.1055/s-0034-1393732.


Evaluation of the rigidity of sagittal split ramus osteotomy fixation using four designs of biodegradable and titanium plates--a numerical study.

Albougha S, Albogha M, Darwich M, Darwich K Oral Maxillofac Surg. 2015; 19(3):281-5.

PMID: 25784154 DOI: 10.1007/s10006-015-0491-8.