» Articles » PMID: 20367092

A Prospective, Multicenter, Randomized-controlled 5-year Study of Hybrid and Fully Etched Implants for the Incidence of Peri-implantitis

Overview
Journal J Periodontol
Date 2010 Apr 7
PMID 20367092
Citations 25
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The dual acid-etched (DAE) implant was commercially introduced in 1996 with a hybrid design incorporating a machined surface in the coronal region from approximately the third thread to the seating surface. This design was intended to reduce the risks of peri-implantitis and other related soft tissue complications that were reported for implants with surface roughness in the coronal region. The objective of this prospective, randomized-controlled clinical trial was to determine the incidence of peri-implantitis for a fully etched implant with the DAE surface extending to the implant platform.

Methods: Patients had implant sites randomly assigned to receive one hybrid control implant and at least one fully etched test implant in support of a short-span fixed restoration to ensure that variables (e.g., demographics, jaw locations, and bone density) were consistent between groups. Prostheses were inserted 2 months after implant placement with follow-up evaluations scheduled annually for 5 years to assess mucosal health based on bleeding on probing, suppuration, and probing depths. Evaluations also included radiographic and mobility assessments.

Results: One hundred twelve patients who were enrolled at seven centers received 139 control and 165 test implants (total: 304 implants). With >5 years of postloading evaluations, there was one declaration of peri-implantitis associated with a control implant that was successfully treated later. Clinical probing and radiographic assessments did not reveal differences between groups in mucosal health outcomes or other signs of peri-implantitis.

Conclusion: Five-year results of this randomized-controlled study showed no increased risk of peri-implantitis for fully etched implants compared to hybrid-designed implants.

Citing Articles

Influence of Implant Surfaces on Peri-Implant Diseases - A Systematic Review and Meta-Analysis.

Hussein A, Shah M, Atieh M, Alhimairi S, Amir-Rad F, Elbishari H Int Dent J. 2024; 75(1):75-85.

PMID: 39532569 PMC: 11806302. DOI: 10.1016/j.identj.2024.10.007.


Zirconia Dental Implant Designs and Surface Modifications: A Narrative Review.

Ciszynski M, Chwaliszewski B, Simka W, Dominiak M, Gedrange T, Hadzik J Materials (Basel). 2024; 17(17).

PMID: 39274592 PMC: 11396535. DOI: 10.3390/ma17174202.


Indian Society of Periodontology Good Clinical Practice Recommendations for Peri-implant Care.

Satpathy A, Grover V, Kumar A, Jain A, Gopalakrishnan D, Grover H J Indian Soc Periodontol. 2024; 28(1):6-31.

PMID: 38988964 PMC: 11232813. DOI: 10.4103/jisp.jisp_124_24.


Comparison of pre- and post-implantation of Indonesian-made plates in fracture patients: Functional, radiological, biomechanical and chemical analyses.

Lisan R, Mahyudin F, Mubarok F Narra J. 2024; 4(1):e752.

PMID: 38798872 PMC: 11125419. DOI: 10.52225/narra.v4i1.752.


Prevalence of Peri-Implant Mucositis, Peri-Implantitis and Associated Risk Indicators of Implants with and without Laser-Microgrooved Collar Surface: A Long-Term (≥20 Years) Retrospective Study.

Guarnieri R, Reda R, Di Nardo D, Pagnoni F, Zanza A, Testarelli L J Pers Med. 2024; 14(4).

PMID: 38672969 PMC: 11050992. DOI: 10.3390/jpm14040342.