The Variable Ellipsoid Modeling Technique As a Verification Method for the Treatment Planning System of Gamma Knife Radiosurgery

Overview
Date 2010 Mar 13
PMID 20224712
Citations 4
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: The secondary verification of Leksell Gamma Knife treatment planning system (LGP) (which is the primary verification system) is extremely important in order to minimize the risk of treatment errors. Although prior methods have been developed to verify maximum dose and treatment time, none have studied maximum dose coordinates and treatment volume.

Methods: We simulated the skull shape as an ellipsoid with its center at the junction between the mammillary bodies and the brain stem. The radiation depths of the beamlets emitted from 201 collimators were calculated based on the relationship between this ellipsoid and a single beamlet expressed as a straight line. A computer program was coded to execute the algorithm. A database system was adopted to log the doses for 31x31x31 or 29,791 matrix points allowing for future queries to be made of the matrix of interest.

Results: When we compared the parameters in seven patients, all parameters showed good correlation. The number of matrix points with a dose higher than 30% of the maximal dose was within +/- 2% of LGP. The 50% dose volume, which is generally the target volume, differs maximally by 4.2%. The difference of the maximal dose ranges from 0.7% to 7%.

Conclusion: Based on the results, the variable ellipsoid modeling technique or variable ellipsoid modeling technique (VEMT) can be a useful and independent tool to verify the important parameters of LGP and make up for LGP.

Citing Articles

Validation of Gamma Knife Perfexion Dose Profile Distribution by a Modified Variable Ellipsoid Modeling Technique.

Hur B, Jin S, Kim G, Kwak J, Kim Y, Lee S J Korean Neurosurg Soc. 2021; 64(1):13-22.

PMID: 33465847 PMC: 7819799. DOI: 10.3340/jkns.2020.0186.


Two-year experience with the commercial Gamma Knife Check software.

Xu A, Bhatnagar J, Bednarz G, Novotny Jr J, Flickinger J, Lunsford L J Appl Clin Med Phys. 2016; 17(4):95-105.

PMID: 27455470 PMC: 5690029. DOI: 10.1120/jacmp.v17i4.5547.


Clinical application of gamma knife dose verification method in multiple brain tumors : modified variable ellipsoid modeling technique.

Hur B, Lee J, Cho W, Kang D, Kim C, Choi B J Korean Neurosurg Soc. 2013; 53(2):102-7.

PMID: 23560174 PMC: 3611052. DOI: 10.3340/jkns.2013.53.2.102.


Radiation-induced intratumoral necrosis and peritumoral edema after gamma knife radiosurgery for intracranial meningiomas.

Lee S, Yang K, Kim S, Kim S J Korean Neurosurg Soc. 2012; 52(2):98-102.

PMID: 23091666 PMC: 3467383. DOI: 10.3340/jkns.2012.52.2.98.

References
1.
SHIELDS J, Medearis A, Bear M . Fetal head and abdominal circumferences: effect of profile shape on the accuracy of ellipse equations. J Clin Ultrasound. 1987; 15(4):241-4. DOI: 10.1002/jcu.1870150404. View

2.
Maitz A, Wu A, Lunsford L, Flickinger J, Kondziolka D, Bloomer W . Quality assurance for gamma knife stereotactic radiosurgery. Int J Radiat Oncol Biol Phys. 1995; 32(5):1465-71. DOI: 10.1016/0360-3016(95)00577-L. View

3.
Marcu S, Wu Q, Pillai K, Weinhous M . GammaPlan-Leksell Gamma Knife radiosurgery treatment planning verification method. Med Phys. 2000; 27(9):2146-9. DOI: 10.1118/1.1287051. View

4.
Jin J, Drzymala R, Li Z . A simple method of independent treatment time verification in gamma knife radiosurgery using integral dose. Med Phys. 2005; 31(12):3337-44. DOI: 10.1118/1.1824192. View

5.
Zhang P, Dean D, Wu Q, Sibata C . Fast verification of Gamma Knifetrade mark treatment plans. J Appl Clin Med Phys. 2001; 1(4):158-64. PMC: 5726158. DOI: 10.1120/jacmp.v1i4.2638. View