Abstract

Background: Reverse shoulder arthroplasty has become more prevalent for the treatment of complex shoulder issues. Prosthetic designs vary in both the number and orientation of screws recommended for securing the glenoid base plate. This study examined the contribution of the posterior glenoid screw for stabilizing the glenosphere by comparing constructs with no posterior screw, a standard posterior screw directed into the glenoid neck, and a long posterior screw (LPS) into the scapular spine.

Material And Methods: The Tornier RTSA glenoid implant was fixed into 2 groups of matched cadaveric scapulae. In both groups, the controls were fixed with a standard posterior screw (SPS). Matching scapulae had a screw configuration that omitted the posterior screw (NPS) in group I or utilized an LPS in group II. Specimens were tested using a "rocking-horse" protocol. During cyclic loading (50,000 cycles), the vertical displacement of the glenoid component was monitored using a digital caliper.

Results: In group I, NPS constructs demonstrated a significantly higher mean rate of loosening than SPS constructs. In group II, the LPS constructs demonstrated lower loosening rates than SPS constructs. Mean initial displacements were greater for NPS than SPS in group I and similar for SPS and LPS in group II DISCUSSION: The posterior glenoid screw contributes significantly to stability of the reverse glenoid base plate. If an SPS does not obtain good purchase into the scapula, glenoid component fixation will be enhanced by the addition of an LPS into the scapular spine.

Citing Articles

Mid-term Results Following Reverse Shoulder Arthroplasty and the Role of Navigation in the Management of Glenoid Bone Loss.

Saraglis G, Singh H, Charfare Z, Olujinmi G, Devecseri G, Agbaje A Cureus. 2024; 16(2):e54633.

PMID: 38524053 PMC: 10959506. DOI: 10.7759/cureus.54633.


Ten technical aspects of baseplate fixation in reverse total shoulder arthroplasty for patients without glenoid bone loss: a systematic review.

Spek R, Hoogervorst L, Brink R, Schoones J, van Deurzen D, van den Bekerom M Clin Shoulder Elb. 2023; 27(1):88-107.

PMID: 38147872 PMC: 10938023. DOI: 10.5397/cise.2023.00493.


Implications of navigation system use for glenoid component placement in reverse shoulder arthroplasty.

Kida H, Urita A, Momma D, Matsui Y, Endo T, Kawamura D Sci Rep. 2022; 12(1):21190.

PMID: 36477208 PMC: 9729232. DOI: 10.1038/s41598-022-25833-8.


Pathogenesis, Evaluation, and Management of Osteolysis After Total Shoulder Arthroplasty.

Kunze K, Krivicich L, Brusalis C, Taylor S, Gulotta L, Dines J Clin Shoulder Elb. 2022; 25(3):244-254.

PMID: 35971608 PMC: 9471816. DOI: 10.5397/cise.2021.00738.


Pre-Operative CT-Based Planning Integrated With Intra-Operative Navigation in Reverse Shoulder Arthroplasty: Data Acquisition and Analysis Protocol, and Preliminary Results of Navigated Versus Conventional Surgery.

Moreschini F, Colasanti G, Cataldi C, Mannelli L, Mondanelli N, Giannotti S Dose Response. 2022; 18(4):1559325820970832.

PMID: 35185413 PMC: 8851139. DOI: 10.1177/1559325820970832.