Estimation of Viscous Dissipative Stresses Induced by a Mechanical Heart Valve Using PIV Data

Overview
Journal Ann Biomed Eng
Date 2009 Dec 19
PMID 20020213
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Among the clinical complications of mechanical heart valves (MHVs), hemolysis was previously thought to result from Reynolds stresses in turbulent flows. A more recent hypothesis suggests viscous dissipative stresses at spatial scales similar in size to red blood cells may be related to hemolysis in MHVs, but the resolution of current instrumentation is insufficient to measure the smallest eddy sizes. We studied the St. Jude Medical (SJM) 27 mm valve in the aortic position of a pulsatile circulatory mock loop under physiologic conditions with particle image velocimetry (PIV). Assuming a dynamic equilibrium assumption between the resolved and sub-grid-scale (SGS) energy flux, the SGS energy flux was calculated from the strain rate tensor computed from the resolved velocity fields and the SGS stress was determined by the Smagorinsky model, from which the turbulence dissipation rate and then the viscous dissipative stresses were estimated. Our results showed Reynolds stresses up to 80 N/m2 throughout the cardiac cycle, and viscous dissipative stresses below 12 N/m2. The viscous dissipative stresses remain far below the threshold of red blood cell hemolysis, but could potentially damage platelets, implying the need for further study in the phenomenon of MHV hemolytic complications.

Citing Articles

Characterization of Turbulent Flow Behind a Transcatheter Aortic Valve in Different Implantation Positions.

Pietrasanta L, Zheng S, De Marinis D, Hasler D, Obrist D Front Cardiovasc Med. 2022; 8:804565.

PMID: 35097022 PMC: 8794584. DOI: 10.3389/fcvm.2021.804565.


Simulated Transcatheter Aortic Valve Flow: Implications of Elliptical Deployment and Under-Expansion at the Aortic Annulus.

Sirois E, Mao W, Li K, Calderan J, Sun W Artif Organs. 2018; 42(7):E141-E152.

PMID: 29608034 PMC: 6097918. DOI: 10.1111/aor.13107.


Assessment of turbulent viscous stress using ICOSA 4D Flow MRI for prediction of hemodynamic blood damage.

Ha H, Lantz J, Haraldsson H, Casas B, Ziegler M, Karlsson M Sci Rep. 2016; 6:39773.

PMID: 28004789 PMC: 5177919. DOI: 10.1038/srep39773.


The effect of turbulent viscous shear stress on red blood cell hemolysis.

Yen J, Chen S, Chern M, Lu P J Artif Organs. 2014; 17(2):178-85.

PMID: 24619800 DOI: 10.1007/s10047-014-0755-3.


Measurements of steady flow through a bileaflet mechanical heart valve using stereoscopic PIV.

Hutchison C, Sullivan P, Ethier C Med Biol Eng Comput. 2010; 49(3):325-35.

PMID: 21080093 DOI: 10.1007/s11517-010-0705-z.