Remote-controlled Scanning and Automated Confocal Microscopy Through Focusing Using a Modified HRT Rostock Corneal Module

Overview
Specialty Ophthalmology
Date 2009 Nov 11
PMID 19901584
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To modify the HRT-II Confocal Microscope with Rostock Corneal Module (HRT-RCM) to allow computerized control of the focal plane position (depth) within the cornea.

Methods: A threaded housing on the HRT-RCM microscope is normally rotated by hand to change the focal plane position within the cornea. This piece was removed to allow the front housing of the microscope to move freely. A linear actuator (Oriel Encoder Mike) was then attached to the side of the microscope and coupled to a drive shaft that was connected to the front housing. The actuator was connected to an Oriel 18011 Encoder Mike controller, which was interfaced to a PC. Software was developed to allow control and display of the focal plane position using this PC, while image acquisition software was run on the HRT-RCM PC. The instrument was tested on one human volunteer.

Results: The modified instrument successfully allowed computer-controlled focusing throughout the entire cornea. Through-focus sequences could be collected online and analyzed and reconstructed three dimensionally off-line using modified confocal microscopy through-focusing software.

Conclusions: Although this is only a prototype instrument, it significantly improves the examination procedure by allowing completely "hands-free" operation of the HRT-RCM microscope. The data also demonstrate the feasibility of performing quantitative z-axis scans through the full thickness of the cornea with the HRT-RCM. Given the higher contrast images and improved optical sectioning of the HRT-RCM as compared with other instruments, these capabilities could have widespread application.

Citing Articles

Loss of Corneal Nerves and Corneal Haze in Patients with Fuchs' Endothelial Corneal Dystrophy with the Transcription Factor 4 Gene Trinucleotide Repeat Expansion.

Gillings M, Mastro A, Zhang X, Kiser K, Gu J, Xing C Ophthalmol Sci. 2022; 3(1):100214.

PMID: 36275201 PMC: 9563205. DOI: 10.1016/j.xops.2022.100214.


Differential effects of obstructive sleep apnea on the corneal subbasal nerve plexus and retinal nerve fiber layer.

Bussan K, Stuard W, Mussi N, Lee W, Whitson J, Issioui Y PLoS One. 2022; 17(6):e0266483.

PMID: 35771778 PMC: 9246161. DOI: 10.1371/journal.pone.0266483.


Wide-Field In Vivo Confocal Microscopy of Meibomian Gland Acini and Rete Ridges in the Eyelid Margin.

Zhou S, Robertson D Invest Ophthalmol Vis Sci. 2018; 59(10):4249-4257.

PMID: 30128496 PMC: 6103323. DOI: 10.1167/iovs.18-24497.


Tear Levels of Insulin-Like Growth Factor Binding Protein 3 Correlate With Subbasal Nerve Plexus Changes in Patients With Type 2 Diabetes Mellitus.

Stuard W, Titone R, Robertson D Invest Ophthalmol Vis Sci. 2017; 58(14):6105-6112.

PMID: 29214310 PMC: 5718599. DOI: 10.1167/iovs.17-22425.


In Vivo Confocal Microscopy of the Cornea: New Developments in Image Acquisition, Reconstruction, and Analysis Using the HRT-Rostock Corneal Module.

Petroll W, Robertson D Ocul Surf. 2015; 13(3):187-203.

PMID: 25998608 PMC: 4499020. DOI: 10.1016/j.jtos.2015.05.002.


References
1.
Vesaluoma M, Petroll W, Linna T, Alio J, Tervo T . Corneal stromal changes induced by myopic LASIK. Invest Ophthalmol Vis Sci. 2000; 41(2):369-76. View

2.
Imayasu M, Petroll W, Jester J, Patel S, Ohashi J, Cavanagh H . The relation between contact lens oxygen transmissibility and binding of Pseudomonas aeruginosa to the cornea after overnight wear. Ophthalmology. 1994; 101(2):371-88. DOI: 10.1016/s0161-6420(94)31326-1. View

3.
Masters B, Thaer A . In vivo human corneal confocal microscopy of identical fields of subepithelial nerve plexus, basal epithelial, and wing cells at different times. Microsc Res Tech. 1994; 29(5):350-6. DOI: 10.1002/jemt.1070290505. View

4.
McCulley J, Petroll W . Quantitative assessment of corneal wound healing following IntraLASIK using in vivo confocal microscopy. Trans Am Ophthalmol Soc. 2009; 106:84-90. PMC: 2646432. View

5.
Wiegand W, Thaer A, Kroll P, GEYER O, Garcia A . Optical sectioning of the cornea with a new confocal in vivo slit-scanning videomicroscope. Ophthalmology. 1995; 102(4):568-75. DOI: 10.1016/s0161-6420(95)30981-5. View