» Articles » PMID: 19662339

Dose Reduction in Spiral CT Coronary Angiography with Dual-source Equipment. Part I. A Phantom Study Applying Different Prospective Tube Current Modulation Algorithms

Overview
Journal Radiol Med
Specialty Radiology
Date 2009 Aug 8
PMID 19662339
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: The authors sought to compare different algorithms for dose reduction in retrospectively echocardiographically (ECG)-gated dual-source computed tomography (CT) coronary angiography (DSCT-CA) in a phantom model.

Materials And Methods: Weighted CT dose index (CTDI) was measured by using an anthropomorphic phantom in spiral cardiac mode (retrospective ECG gating) at five pitch values adapted with two heart-rate-adaptive ECG pulsing windows using four algorithms: narrow pulsing window, with tube current reduction to 20% (A) and 4% (B) of peak current outside the pulsing window; wide pulsing window, with tube current reduction to 20% (C) and 4% (D). Each algorithm was applied at different heart rates (45, 60, 75, 90, 120 bpm).

Results: Mean CTDI volume (CTDIvol) was 36.9+/-9.7 mGy, 23.9+/-5.6 mGy, 49.7+/-16.2 mGy and 38.5+/-12.3 mGy for A, B, C and D, respectively. Consistent dose reduction was observed with protocols applying the 4% tube current reduction (B and D). Using the conversion coefficient for the chest, the mean effective dose was the highest for C (9.6 mSv) and the lowest for B (4.6 mSv). Heart-ratedependent pitch values (pitch=0.2, 0.26, 0.34, 0.43, 0.5) and the use of heart-rate-adaptive ECG pulsing windows provided a significant decrease in the CTDIvol with progressively higher heart rates (45, 60, 75, 90, 120 bpm), despite using wider pulsing windows.

Conclusions: Radiation exposure with DSCT-CA using a narrow pulsing window significantly decreases when compared with a wider pulsing window. When using a protocol with reduced tube current to 4%, the radiation dose is significantly lower.

Citing Articles

256-slice CT coronary angiography: in vivo dosimetry and technique optimization.

Belgrano M, Bregant P, Djoguela M, Toscano W, Marchese E, Cova M Radiol Med. 2013; 119(4):249-56.

PMID: 24337756 DOI: 10.1007/s11547-013-0334-3.


The Role of Cardiovascular Magnetic Resonance (CMR) and Computed Tomography (CCT) in Facilitating Heart Failure Management.

Stacey R, Hundley W Curr Treat Options Cardiovasc Med. 2013; 15(4):373-86.

PMID: 23817725 PMC: 3767383. DOI: 10.1007/s11936-013-0253-6.


Prevalence of Congenital Coronary Artery Anomalies and Variants in 2697 Consecutive Patients Using 64-Detector Row Coronary CTAngiography.

Shabestari A, Akhlaghpoor S, Tayebivaljozi R, Fattahi Masrour F Iran J Radiol. 2013; 9(3):111-21.

PMID: 23329976 PMC: 3522371. DOI: 10.5812/iranjradiol.8070.


Diagnostic accuracy of 64-slice computed tomography coronary angiography in a large population of patients without revascularisation: registry data on the comparison between male and female population.

Maffei E, Martini C, Tedeschi C, Spagnolo P, Zuccarelli A, Arcadi T Radiol Med. 2011; 117(1):6-18.

PMID: 21643636 DOI: 10.1007/s11547-011-0693-6.


Computed tomography coronary angiography in the selection of outlier patients: a feasibility report.

Maffei E, Martini C, Seitun S, Arcadi T, Tedeschi C, Guaricci A Radiol Med. 2011; 117(2):214-29.

PMID: 21643634 DOI: 10.1007/s11547-011-0695-4.


References
1.
Cademartiri F, Luccichenti G, Marano R, Nieman K, Mollet N, de Feyter P . Non-invasive angiography of the coronary arteries with multislice computed tomography: state of the art and future prospects. Radiol Med. 2003; 106(4):284-96. View

2.
Ropers U, Ropers D, Pflederer T, Anders K, Kuettner A, Stilianakis N . Influence of heart rate on the diagnostic accuracy of dual-source computed tomography coronary angiography. J Am Coll Cardiol. 2007; 50(25):2393-8. DOI: 10.1016/j.jacc.2007.09.017. View

3.
Giesler T, Baum U, Ropers D, Ulzheimer S, Wenkel E, Mennicke M . Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT: influence of heart rate on image quality and stenosis detection. AJR Am J Roentgenol. 2002; 179(4):911-6. DOI: 10.2214/ajr.179.4.1790911. View

4.
Flohr T, McCollough C, Bruder H, Petersilka M, Gruber K, Suss C . First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol. 2005; 16(2):256-68. DOI: 10.1007/s00330-005-2919-2. View

5.
Passariello R, De Santis M . [Coronary artery disease. Update and prospects of radiologic imaging with CT and MR]. Radiol Med. 2001; 101(6):411-23. View