Dose Reduction in Spiral CT Coronary Angiography with Dual-source Equipment. Part I. A Phantom Study Applying Different Prospective Tube Current Modulation Algorithms
Overview
Authors
Affiliations
Purpose: The authors sought to compare different algorithms for dose reduction in retrospectively echocardiographically (ECG)-gated dual-source computed tomography (CT) coronary angiography (DSCT-CA) in a phantom model.
Materials And Methods: Weighted CT dose index (CTDI) was measured by using an anthropomorphic phantom in spiral cardiac mode (retrospective ECG gating) at five pitch values adapted with two heart-rate-adaptive ECG pulsing windows using four algorithms: narrow pulsing window, with tube current reduction to 20% (A) and 4% (B) of peak current outside the pulsing window; wide pulsing window, with tube current reduction to 20% (C) and 4% (D). Each algorithm was applied at different heart rates (45, 60, 75, 90, 120 bpm).
Results: Mean CTDI volume (CTDIvol) was 36.9+/-9.7 mGy, 23.9+/-5.6 mGy, 49.7+/-16.2 mGy and 38.5+/-12.3 mGy for A, B, C and D, respectively. Consistent dose reduction was observed with protocols applying the 4% tube current reduction (B and D). Using the conversion coefficient for the chest, the mean effective dose was the highest for C (9.6 mSv) and the lowest for B (4.6 mSv). Heart-ratedependent pitch values (pitch=0.2, 0.26, 0.34, 0.43, 0.5) and the use of heart-rate-adaptive ECG pulsing windows provided a significant decrease in the CTDIvol with progressively higher heart rates (45, 60, 75, 90, 120 bpm), despite using wider pulsing windows.
Conclusions: Radiation exposure with DSCT-CA using a narrow pulsing window significantly decreases when compared with a wider pulsing window. When using a protocol with reduced tube current to 4%, the radiation dose is significantly lower.
256-slice CT coronary angiography: in vivo dosimetry and technique optimization.
Belgrano M, Bregant P, Djoguela M, Toscano W, Marchese E, Cova M Radiol Med. 2013; 119(4):249-56.
PMID: 24337756 DOI: 10.1007/s11547-013-0334-3.
Stacey R, Hundley W Curr Treat Options Cardiovasc Med. 2013; 15(4):373-86.
PMID: 23817725 PMC: 3767383. DOI: 10.1007/s11936-013-0253-6.
Shabestari A, Akhlaghpoor S, Tayebivaljozi R, Fattahi Masrour F Iran J Radiol. 2013; 9(3):111-21.
PMID: 23329976 PMC: 3522371. DOI: 10.5812/iranjradiol.8070.
Maffei E, Martini C, Tedeschi C, Spagnolo P, Zuccarelli A, Arcadi T Radiol Med. 2011; 117(1):6-18.
PMID: 21643636 DOI: 10.1007/s11547-011-0693-6.
Computed tomography coronary angiography in the selection of outlier patients: a feasibility report.
Maffei E, Martini C, Seitun S, Arcadi T, Tedeschi C, Guaricci A Radiol Med. 2011; 117(2):214-29.
PMID: 21643634 DOI: 10.1007/s11547-011-0695-4.