Abstract

Design: Cross-sectional studies in Russia (n = 502) and Macedonia (n = 205), with fluorochrome-stained sputum examined by 1) the new Lumin light emitting diode (LED) fluorescent attachment on a light microscope, and 2) conventional fluorescent microscope (CFM) available in each laboratory, and compared to 3) Ziehl-Neelsen (ZN) restaining/reading of the same smears. Poor readings of ZN-restained smears in Russia stimulated a retrospective laboratory registry analysis for sensitivity and specificity of directly ZN-stained smears (n = 791) from a previous period.

Results: In Macedonia, the sensitivity of the Lumin and CFM were 87.8%, and that of restained ZN smears with conventional light microscope was 78.0%. In Russia, sensitivity was as follows: Lumin 72.8%, CFM 52.5%; re-stained ZN smears 28.5% and directly ZN stained smears 55.6%.

Conclusion: Fluorescence microscopy is more sensitive than conventional microscopy. The Lumin attachment to conventional light microscopes provided results equal to or better than the CFMs. Smear restaining for ZN showed a 12% advantage for Lumin and CFM in Macedonia, in line with other meta-analyses. Restaining for ZN gave poor results in Russia for unknown reasons. Retrospective analysis of directly ZN-stained smears showed 55.6% sensitivity compared to the Lumin (72.8%), which is also in line with the superiority of fluorescent microscopy reported in literature.

Citing Articles

Evaluating a Dual-Gene qPCR Melting Curve Assay for Rapid Detection of Tuberculosis in Suspected Pulmonary Cases.

Ma J, Wang L, Zheng X, Cui H, Sha W Infect Drug Resist. 2025; 18:1135-1147.

PMID: 40027917 PMC: 11871878. DOI: 10.2147/IDR.S498180.


Evaluation of modified bleach technique for the detection of acid fast bacilli in lymph node aspirate at the University of Gondar Comprehensive Specialized Hospital, Northwest Ethiopia.

Mulugeta F, Tiruneh M, Abebe B, Yitayew G, Ayehubizu Z, Getahun M J Clin Tuberc Other Mycobact Dis. 2022; 28:100328.

PMID: 35990771 PMC: 9386082. DOI: 10.1016/j.jctube.2022.100328.


Ozonation enhancement of low cost double-stranded DNA binding dye based fluorescence measurement of total bacterial load in water.

Choi J, Chua B, Son A RSC Adv. 2022; 11(7):3931-3941.

PMID: 35424342 PMC: 8694141. DOI: 10.1039/d0ra08742d.


Diagnostic performance of smear microscopy and incremental yield of Xpert in detection of pulmonary tuberculosis in Rwanda.

Ngabonziza J, Ssengooba W, Mutua F, Torrea G, Dushime A, Gasana M BMC Infect Dis. 2016; 16(1):660.

PMID: 27825314 PMC: 5101805. DOI: 10.1186/s12879-016-2009-x.


Rapid culture-based diagnosis of pulmonary tuberculosis in developed and developing countries.

Asmar S, Drancourt M Front Microbiol. 2015; 6:1184.

PMID: 26579092 PMC: 4630581. DOI: 10.3389/fmicb.2015.01184.