Purpose: To assess patient setup corrections based on daily megavoltage CT (MVCT) imaging for four anatomic treatment sites treated on tomotherapy.

Method And Materials: Translational and rotational setup corrections, based on registration of daily MVCT to planning CT images, were analyzed for 1,179 brain and head and neck (H&N), 1,414 lung, and 1,274 prostate treatment fractions. Frequencies of three-dimensional vector lengths, overall distributions of setup corrections, and patient-specific distributions of random and systematic setup errors were analyzed.

Results: Brain and H&N had lower magnitude positioning corrections and smaller variations in translational setup errors but were comparable in roll rotations. Three-dimensional vector translational shifts of larger magnitudes occurred more frequently for lung and prostate than for brain and H&N treatments, yet this was not observed for roll rotations. The global systematic error for prostate was 4.7 mm in the vertical direction, most likely due to couch sag caused by large couch extension distances. Variations in systematic errors and magnitudes of random translational errors ranged from 1.6 to 2.6 mm for brain and H&N and 3.2 to 7.2 mm for lung and prostate, whereas roll rotational errors ranged from 0.8 degrees to 1.2 degrees for brain and H&N and 0.5 degrees to 1.0 degrees for lung and prostate.

Conclusions: Differences in setup were observed between brain, H&N, lung, and prostate treatments. Patient setup can be improved if daily imaging is performed. This analysis can assess the utilization of daily image guidance and allows for further investigation into improved anatomic site-specific and patient-specific treatments.

Citing Articles

The top 100 most cited articles in helical tomotherapy: a scoping review.

Zhang Y, Rong L, Wang Z, Zhao H Front Oncol. 2023; 13:1274290.

PMID: 37916164 PMC: 10616822. DOI: 10.3389/fonc.2023.1274290.


Feasibility of artificial intelligence-driven interfractional monitoring of organ changes by mega-voltage computed tomography in intensity-modulated radiotherapy of prostate cancer.

Lee Y, Choi H, Kim H, Kim S, Kim M, Cha H Radiat Oncol J. 2023; 41(3):186-198.

PMID: 37793628 PMC: 10556843. DOI: 10.3857/roj.2023.00444.


Assessing setup errors and shifting margins for planning target volume in head, neck, and breast cancer.

Alabedi H J Med Life. 2023; 16(3):394-398.

PMID: 37168304 PMC: 10165517. DOI: 10.25122/jml-2022-0241.


Setup Accuracy in Craniospinal Irradiation: Implications for Planning Treatment Volume Margins.

Novak J, Du D, Shinde A, Li R, Amini A, Chen Y Adv Radiat Oncol. 2021; 6(5):100747.

PMID: 34646966 PMC: 8498716. DOI: 10.1016/j.adro.2021.100747.


Surface-guided tomotherapy improves positioning and reduces treatment time: A retrospective analysis of 16 835 treatment fractions.

Haraldsson A, Ceberg S, Ceberg C, Back S, Engelholm S, Engstrom P J Appl Clin Med Phys. 2020; 21(8):139-148.

PMID: 32592288 PMC: 7484821. DOI: 10.1002/acm2.12936.


References
1.
Mackie T . History of tomotherapy. Phys Med Biol. 2006; 51(13):R427-53. DOI: 10.1088/0031-9155/51/13/R24. View

2.
Yan D, Lockman D, Martinez A, Wong J, Brabbins D, Vicini F . Computed tomography guided management of interfractional patient variation. Semin Radiat Oncol. 2005; 15(3):168-79. DOI: 10.1016/j.semradonc.2005.01.007. View

3.
Guckenberger M, Meyer J, Vordermark D, Baier K, Wilbert J, Flentje M . Magnitude and clinical relevance of translational and rotational patient setup errors: a cone-beam CT study. Int J Radiat Oncol Biol Phys. 2006; 65(3):934-42. DOI: 10.1016/j.ijrobp.2006.02.019. View

4.
Mackie T, Balog J, Ruchala K, Shepard D, Aldridge S, Fitchard E . Tomotherapy. Semin Radiat Oncol. 1999; 9(1):108-17. DOI: 10.1016/s1053-4296(99)80058-7. View

5.
Boswell S, Tome W, Jeraj R, Jaradat H, Mackie T . Automatic registration of megavoltage to kilovoltage CT images in helical tomotherapy: an evaluation of the setup verification process for the special case of a rigid head phantom. Med Phys. 2006; 33(11):4395-404. DOI: 10.1118/1.2349698. View