Image Quality, Tissue Heating, and Frame Rate Trade-offs in Acoustic Radiation Force Impulse Imaging

Overview
Authors
Affiliations
Soon will be listed here.
Abstract

The real-time application of acoustic radiation force impulse (ARFI) imaging requires both short acquisition times for a single ARFI image and repeated acquisition of these frames. Due to the high energy of pulses required to generate appreciable radiation force, however, repeated acquisitions could result in substantial transducer face and tissue heating. We describe and evaluate several novel beam sequencing schemes which, along with parallel-receive acquisition, are designed to reduce acquisition time and heating. These techniques reduce the total number of radiation force impulses needed to generate an image and minimize the time between successive impulses. We present qualitative and quantitative analyses of the trade-offs in image quality resulting from the acquisition schemes. Results indicate that these techniques yield a significant improvement in frame rate with only moderate decreases in image quality. Tissue and transducer face heating resulting from these schemes is assessed through finite element method modeling and thermocouple measurements. Results indicate that heating issues can be mitigated by employing ARFI acquisition sequences that utilize the highest track-to-excitation ratio possible.

Citing Articles

Shear wave elastography primer for the abdominal radiologist.

Urban M, Vasconcelos L, Brom K, Dave J, Kijanka P Abdom Radiol (NY). 2025; .

PMID: 39883164 DOI: 10.1007/s00261-025-04806-1.


Projection-based stereolithography for direct 3D printing of heterogeneous ultrasound phantoms.

Paulsen S, Mitcham T, Pan C, Long J, Grigoryan B, Sazer D PLoS One. 2021; 16(12):e0260737.

PMID: 34882719 PMC: 8659365. DOI: 10.1371/journal.pone.0260737.


Synchronous temperature variation monitoring during ultrasound imaging and/or treatment pulse application: a phantom study.

Kamimura H, Saharkhiz N, Lee S, Konofagou E IEEE Open J Ultrason Ferroelectr Freq Control. 2021; 1:1-10.

PMID: 34713274 PMC: 8547607. DOI: 10.1109/ojuffc.2021.3085539.


Hydrophone Spatial Averaging Correction for Acoustic Exposure Measurements From Arrays-Part I: Theory and Impact on Diagnostic Safety Indexes.

Wear K IEEE Trans Ultrason Ferroelectr Freq Control. 2020; 68(3):358-375.

PMID: 33186102 PMC: 8325172. DOI: 10.1109/TUFFC.2020.3037946.


Elastographic Tomosynthesis From X-Ray Strain Imaging of Breast Cancer.

Sutphin C, Olson E, Motai Y, Lee S, Kim J, Takabe K IEEE J Transl Eng Health Med. 2019; 7:4300312.

PMID: 31497411 PMC: 6726464. DOI: 10.1109/JTEHM.2019.2935721.


References
1.
Palmeri M, McAleavey S, Fong K, Trahey G, Nightingale K . Dynamic mechanical response of elastic spherical inclusions to impulsive acoustic radiation force excitation. IEEE Trans Ultrason Ferroelectr Freq Control. 2006; 53(11):2065-79. PMC: 1815395. DOI: 10.1109/tuffc.2006.146. View

2.
Konofagou E, Ottensmeyer M, Agabian S, Dawson S, Hynynen K . Estimating localized oscillatory tissue motion for assessment of the underlying mechanical modulus. Ultrasonics. 2004; 42(1-9):951-6. DOI: 10.1016/j.ultras.2003.12.001. View

3.
Palmeri M, Nightingale K . On the thermal effects associated with radiation force imaging of soft tissue. IEEE Trans Ultrason Ferroelectr Freq Control. 2004; 51(5):551-65. View

4.
Fahey B, Palmeri M, Trahey G . The impact of physiological motion on tissue tracking during radiation force imaging. Ultrasound Med Biol. 2007; 33(7):1149-66. PMC: 2075097. DOI: 10.1016/j.ultrasmedbio.2007.01.007. View

5.
Dahl J, Pinton G, Palmeri M, Agrawal V, Nightingale K, Trahey G . A parallel tracking method for acoustic radiation force impulse imaging. IEEE Trans Ultrason Ferroelectr Freq Control. 2007; 54(2):301-12. PMC: 1810393. DOI: 10.1109/tuffc.2007.244. View