» Articles » PMID: 19130239

Sample Preparation Module for Bacterial Lysis and Isolation of DNA from Human Urine

Overview
Date 2009 Jan 9
PMID 19130239
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Silica impregnated polymer monolithic columns may provide a simple method for lysing and extracting DNA from bacteria inside of microfluidic chips. Here we use Escherichia coli as a test organism for a point of care thermoplastic microfluidic module designed to take in a urine sample, mix it with lysis buffer, and perform a hybrid chemical/mechanical lysis and solid phase extraction of nucleic acids from the sample. To demonstrate proof-of-concept, we doped human hematuric urine samples with E. coli at concentrations ranging from 10(1)-10(5) colony-forming units/mL (CFU/mL) to simulate patient samples. We then performed on-chip lysis and DNA extraction. The bacterial DNA was amplified using real-time PCR demonstrating lysis and isolation down to 10(1) CFU/mL. Results were comparable to a commercial kit at higher concentrations and performed better at recovering DNA at lower concentrations.

Citing Articles

Sample preparation and detection methods in point-of-care devices towards future at-home testing.

Adedokun G, Alipanah M, Fan Z Lab Chip. 2024; 24(15):3626-3650.

PMID: 38952234 PMC: 11270053. DOI: 10.1039/d3lc00943b.


MC-PRPA-HLFIA Cascade Detection System for Point-of-Care Testing Pan-Drug-Resistant Genes in Urinary Tract Infection Samples.

Tao J, Liu D, Xiong J, Shan W, Dou L, Zhai W Int J Mol Sci. 2023; 24(7).

PMID: 37047757 PMC: 10095522. DOI: 10.3390/ijms24076784.


An Overview on Microfluidic Systems for Nucleic Acids Extraction from Human Raw Samples.

Obino D, Vassalli M, Franceschi A, Alessandrini A, Facci P, Viti F Sensors (Basel). 2021; 21(9).

PMID: 33925730 PMC: 8125272. DOI: 10.3390/s21093058.


Advances in point-of-care nucleic acid extraction technologies for rapid diagnosis of human and plant diseases.

Paul R, Ostermann E, Wei Q Biosens Bioelectron. 2020; 169:112592.

PMID: 32942143 PMC: 7476893. DOI: 10.1016/j.bios.2020.112592.


A paperfluidic platform to detect Neisseria gonorrhoeae in clinical samples.

Horst A, Rosenbohm J, Kolluri N, Hardick J, Gaydos C, Cabodi M Biomed Microdevices. 2018; 20(2):35.

PMID: 29644437 PMC: 6154386. DOI: 10.1007/s10544-018-0280-x.


References
1.
Breadmore M, Shrinivasan S, Wolfe K, Power M, Ferrance J, Hosticka B . Towards a microchip-based chromatographic platform. Part 1: Evaluation of sol-gel phases for capillary electrochromatography. Electrophoresis. 2002; 23(20):3487-95. DOI: 10.1002/1522-2683(200210)23:20<3487::AID-ELPS3487>3.0.CO;2-5. View

2.
Van Dilla M, Langlois R, Pinkel D, Yajko D, Hadley W . Bacterial characterization by flow cytometry. Science. 1983; 220(4597):620-2. DOI: 10.1126/science.6188215. View

3.
Brown P, Freeman A, Foxman B . Prevalence and predictors of trimethoprim-sulfamethoxazole resistance among uropathogenic Escherichia coli isolates in Michigan. Clin Infect Dis. 2002; 34(8):1061-6. DOI: 10.1086/339491. View

4.
Easley C, Karlinsey J, Bienvenue J, Legendre L, Roper M, Feldman S . A fully integrated microfluidic genetic analysis system with sample-in-answer-out capability. Proc Natl Acad Sci U S A. 2006; 103(51):19272-7. PMC: 1748216. DOI: 10.1073/pnas.0604663103. View

5.
Di Carlo D, Ionescu-Zanetti C, Zhang Y, Hung P, Lee L . On-chip cell lysis by local hydroxide generation. Lab Chip. 2005; 5(2):171-8. DOI: 10.1039/b413139h. View