Clinical Characterization of a Proton Beam Continuous Uniform Scanning System with Dose Layer Stacking

Overview
Journal Med Phys
Specialty Biophysics
Date 2008 Dec 17
PMID 19070228
Citations 23
Authors
Affiliations
Soon will be listed here.
Abstract

A proton beam delivery system on a gantry with continuous uniform scanning and dose layer stacking at the Midwest Proton Radiotherapy Institute has been commissioned and accepted for clinical use. This paper was motivated by a lack of guidance on the testing and characterization for clinical uniform scanning systems. As such, it describes how these tasks were performed with a uniform scanning beam delivery system. This paper reports the methods used and important dosimetric characteristics of radiation fields produced by the system. The commissioning data include the transverse and longitudinal dose distributions, penumbra, and absolute dose values. Using a 208 MeV cyclotron's proton beam, the system provides field sizes up to 20 and 30 cm in diameter for proton ranges in water up to 27 and 20 cm, respectively. The dose layer stacking method allows for the flexible construction of spread-out Bragg peaks with uniform modulation of up to 15 cm in water, at typical dose rates of 1-3 Gy/min. For measuring relative dose distributions, multielement ion chamber arrays, small-volume ion chambers, and radiographic films were employed. Measurements during the clinical commissioning of the system have shown that the lateral and longitudinal dose uniformity of 2.5% or better can be achieved for all clinically important field sizes and ranges. The measured transverse penumbra widths offer a slight improvement in comparison to those achieved with a double scattering beam spreading technique at the facility. Absolute dose measurements were done using calibrated ion chambers, thermoluminescent and alanine detectors. Dose intercomparisons conducted using various types of detectors traceable to a national standards laboratory indicate that the measured dosimetry data agree with each other within 5%.

Citing Articles

Validation of pencil beam scanning proton therapy with multi-leaf collimator calculated by a commercial Monte Carlo dose engine.

Tominaga Y, Sakurai Y, Miyata J, Harada S, Akagi T, Oita M J Appl Clin Med Phys. 2022; 23(12):e13817.

PMID: 36420959 PMC: 9797166. DOI: 10.1002/acm2.13817.


Development of Ultra-High Dose-Rate (FLASH) Particle Therapy.

Kim M, Darafsheh A, Schuemann J, Dokic I, Lundh O, Zhao T IEEE Trans Radiat Plasma Med Sci. 2022; 6(3):252-262.

PMID: 36092270 PMC: 9457346. DOI: 10.1109/trpms.2021.3091406.


Mechanical Characterization and Validation of the Dynamic Collimation System Prototype for Proton Radiotherapy.

Geoghegan T, Patwardhan K, Nelson N, Hill P, Flynn R, Smith B J Med Device. 2022; 16(2):021013.

PMID: 35284033 PMC: 8905094. DOI: 10.1115/1.4053722.


Comprehensive Output Estimation of Double Scattering Proton System With Analytical and Machine Learning Models.

Zhu J, Cui T, Zhang Y, Zhang Y, Ma C, Liu B Front Oncol. 2022; 11:756503.

PMID: 35174065 PMC: 8841866. DOI: 10.3389/fonc.2021.756503.


AAPM Task Group Report 290: Respiratory motion management for particle therapy.

Li H, Dong L, Bert C, Chang J, Flampouri S, Jee K Med Phys. 2022; 49(4):e50-e81.

PMID: 35066871 PMC: 9306777. DOI: 10.1002/mp.15470.


References
1.
Anferov V . Combined X-Y scanning magnet for conformal proton radiation therapy. Med Phys. 2005; 32(3):815-8. DOI: 10.1118/1.1862801. View

2.
KIRBY T, Hanson W, Gastorf R, Chu C, SHALEK R . Mailable TLD system for photon and electron therapy beams. Int J Radiat Oncol Biol Phys. 1986; 12(2):261-5. DOI: 10.1016/0360-3016(86)90107-0. View

3.
Gardey K, Oelfke U, Lam G . Range modulation in proton therapy--an optimization technique for clinical and experimental applications. Phys Med Biol. 1999; 44(6):N81-8. DOI: 10.1088/0031-9155/44/6/402. View

4.
Kanai T, Kawachi K, Kumamoto Y, Ogawa H, Yamada T, Matsuzawa H . Spot scanning system for proton radiotherapy. Med Phys. 1980; 7(4):365-9. DOI: 10.1118/1.594693. View

5.
Arduini G, Cambria R, Canzi C, Gerardi F, Gottschalk B, Leone R . Physical specifications of clinical proton beams from a synchrotron. Med Phys. 1996; 23(6):939-51. DOI: 10.1118/1.597735. View