» Articles » PMID: 18937016

Ultrasound Monitoring of a Novel Microwave Ablation (MWA) Device in Porcine Liver: Lessons Learned and Phenomena Observed on Ablative Effects Near Major Intrahepatic Vessels

Overview
Specialty Gastroenterology
Date 2008 Oct 22
PMID 18937016
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Background: Microwave ablation (MWA) is postulated to have several advantages over other thermoablative modalities in the treatment of hepatic tumors. Herein, we use an in vivo porcine model to determine the effect of hepatic blood flow on a novel MWA applicator.

Methods: Four 100-kg pigs underwent hepatic MWA (2,450 MHz, 100 W, 4 min) using a 5.7-mm diameter applicator (Microsulis Americas, Sulis V) inserted near large intrahepatic blood vessels. Real-time monitoring was performed using 3, 5, and 12 MHz diagnostic ultrasound transducers. The ablated zones were sectioned for gross and histological processing.

Results: Ablation zones were uniform in shape and size (3-4 cm) and related to power deliver only. Gross and microscopic examination revealed direct extension of ablation zones to the margin of major hepatic blood vessels and occasionally beyond the intended target. Of note, a momentary acoustic white-out occurred around the probe at 25 +/- -1 s in every ablation.

Discussion: The Sulis V MWA applicator produced uniform zones of ablation that remain unaffected by convective heat loss. The applicator induced a reproducible but temporary event as seen by ultrasound. Further study is warranted to define the physics, benefits, limits, and clinical safety of this new MWA technology.

Citing Articles

Effective and Safe Treatment of Risky Localized Liver Tumors Using Microwave Ablation in Patients with Comorbid Lung Diseases Unfit for Deep Anesthesia.

Asik M Medeni Med J. 2024; 39(1):8-15.

PMID: 38511680 PMC: 10961661. DOI: 10.4274/MMJ.galenos.2024.54358.


Techniques for Temperature Monitoring of Myocardial Tissue Undergoing Radiofrequency Ablation Treatments: An Overview.

Zaltieri M, Massaroni C, Cauti F, Schena E Sensors (Basel). 2021; 21(4).

PMID: 33669692 PMC: 7922285. DOI: 10.3390/s21041453.


Monitoring Thermal Ablation via Microwave Tomography: An Ex Vivo Experimental Assessment.

Scapaticci R, Lopresto V, Pinto R, Cavagnaro M, Crocco L Diagnostics (Basel). 2018; 8(4).

PMID: 30563280 PMC: 6316129. DOI: 10.3390/diagnostics8040081.


Liver microwave ablation: a systematic review of various FDA-approved systems.

Ruiter S, Heerink W, de Jong K Eur Radiol. 2018; 29(8):4026-4035.

PMID: 30506218 PMC: 6611060. DOI: 10.1007/s00330-018-5842-z.


Feasibility of Using a Novel 2.45 GHz Double Short Distance Slot Coaxial Antenna for Minimally Invasive Cancer Breast Microwave Ablation Therapy: Computational Model, Phantom, and Swine Experimentation.

Ortega-Palacios R, Trujillo-Romero C, Cepeda Rubio M, Vera A, Leija L, Reyes J J Healthc Eng. 2018; 2018:5806753.

PMID: 29854360 PMC: 5964617. DOI: 10.1155/2018/5806753.


References
1.
Sato M, Watanabe Y, Ueda S, Iseki S, Abe Y, Sato N . Microwave coagulation therapy for hepatocellular carcinoma. Gastroenterology. 1996; 110(5):1507-14. DOI: 10.1053/gast.1996.v110.pm8613057. View

2.
Chinn S, Lee Jr F, Kennedy G, Chinn C, Johnson C, Winter 3rd T . Effect of vascular occlusion on radiofrequency ablation of the liver: results in a porcine model. AJR Am J Roentgenol. 2001; 176(3):789-95. DOI: 10.2214/ajr.176.3.1760789. View

3.
Haemmerich D, Lee Jr F . Multiple applicator approaches for radiofrequency and microwave ablation. Int J Hyperthermia. 2005; 21(2):93-106. DOI: 10.1080/02656730412331286894. View

4.
Wright A, Sampson L, Warner T, Mahvi D, Lee Jr F . Radiofrequency versus microwave ablation in a hepatic porcine model. Radiology. 2005; 236(1):132-9. DOI: 10.1148/radiol.2361031249. View

5.