Abstract

Background: Based upon defining a common reference point, current real-time quantitative PCR technologies compare relative differences in amplification profile position. As such, absolute quantification requires construction of target-specific standard curves that are highly resource intensive and prone to introducing quantitative errors. Sigmoidal modeling using nonlinear regression has previously demonstrated that absolute quantification can be accomplished without standard curves; however, quantitative errors caused by distortions within the plateau phase have impeded effective implementation of this alternative approach.

Results: Recognition that amplification rate is linearly correlated to amplicon quantity led to the derivation of two sigmoid functions that allow target quantification via linear regression analysis. In addition to circumventing quantitative errors produced by plateau distortions, this approach allows the amplification efficiency within individual amplification reactions to be determined. Absolute quantification is accomplished by first converting individual fluorescence readings into target quantity expressed in fluorescence units, followed by conversion into the number of target molecules via optical calibration. Founded upon expressing reaction fluorescence in relation to amplicon DNA mass, a seminal element of this study was to implement optical calibration using lambda gDNA as a universal quantitative standard. Not only does this eliminate the need to prepare target-specific quantitative standards, it relegates establishment of quantitative scale to a single, highly defined entity. The quantitative competency of this approach was assessed by exploiting "limiting dilution assay" for absolute quantification, which provided an independent gold standard from which to verify quantitative accuracy. This yielded substantive corroborating evidence that absolute accuracies of +/- 25% can be routinely achieved. Comparison with the LinReg and Miner automated qPCR data processing packages further demonstrated the superior performance of this kinetic-based methodology.

Conclusion: Called "linear regression of efficiency" or LRE, this novel kinetic approach confers the ability to conduct high-capacity absolute quantification with unprecedented quality control capabilities. The computational simplicity and recursive nature of LRE quantification also makes it amenable to software implementation, as demonstrated by a prototypic Java program that automates data analysis. This in turn introduces the prospect of conducting absolute quantification with little additional effort beyond that required for the preparation of the amplification reactions.

Citing Articles

Normalizing real-time PCR results in routine testing.

Armenta-Leyva B, Munguia-Ramirez B, Cheng T, Ye F, Henao-Diaz A, Gimenez-Lirola L J Vet Diagn Invest. 2023; 36(1):78-85.

PMID: 37919959 PMC: 10734596. DOI: 10.1177/10406387231206080.


Realizing the value in "non-standard" parts of the qPCR standard curve by integrating fundamentals of quantitative microbiology.

Schmidt P, Acosta N, Chik A, DAoust P, Delatolla R, Dhiyebi H Front Microbiol. 2023; 14:1048661.

PMID: 36937263 PMC: 10020645. DOI: 10.3389/fmicb.2023.1048661.


Effect of on Follicle Development of Djungarian Hamster () with the Variation of Ambient Temperatures.

Qi Y, Xue H, Xu J, Wu M, Chen L, Xu L Biology (Basel). 2023; 12(2).

PMID: 36829590 PMC: 9953326. DOI: 10.3390/biology12020315.


Adaptive Filtering Framework to Remove Nonspecific and Low-Efficiency Reactions in Multiplex Digital PCR Based on Sigmoidal Trends.

Miglietta L, Xu K, Chhaya P, Kreitmann L, Hill-Cawthorne K, Bolt F Anal Chem. 2022; 94(41):14159-14168.

PMID: 36190816 PMC: 9583074. DOI: 10.1021/acs.analchem.2c01883.


Evaluation validation of a qPCR curve analysis method and conventional approaches.

Zhang Y, Li H, Shang S, Meng S, Lin T, Zhang Y BMC Genomics. 2021; 22(Suppl 5):680.

PMID: 34789146 PMC: 8596907. DOI: 10.1186/s12864-021-07986-4.


References
1.
Fan C, Oh D, Wessels L, Weigelt B, Nuyten D, Nobel A . Concordance among gene-expression-based predictors for breast cancer. N Engl J Med. 2006; 355(6):560-9. DOI: 10.1056/NEJMoa052933. View

2.
Skern R, Frost P, Nilsen F . Relative transcript quantification by quantitative PCR: roughly right or precisely wrong?. BMC Mol Biol. 2005; 6:10. PMC: 1090581. DOI: 10.1186/1471-2199-6-10. View

3.
Wang Y, Barbacioru C, Hyland F, Xiao W, Hunkapiller K, Blake J . Large scale real-time PCR validation on gene expression measurements from two commercial long-oligonucleotide microarrays. BMC Genomics. 2006; 7:59. PMC: 1435885. DOI: 10.1186/1471-2164-7-59. View

4.
Gentle A, Anastasopoulos F, McBrien N . High-resolution semi-quantitative real-time PCR without the use of a standard curve. Biotechniques. 2001; 31(3):502, 504-6, 508. DOI: 10.2144/01313st03. View

5.
Chervoneva I, Li Y, Iglewicz B, Waldman S, Hyslop T . Relative quantification based on logistic models for individual polymerase chain reactions. Stat Med. 2007; 26(30):5596-611. DOI: 10.1002/sim.3127. View