Multiplexed Continuous-wave Diode-laser Cavity Ringdown Measurements of Multiple Species
Overview
Overview
Authors
Affiliations
Affiliations
Soon will be listed here.
Abstract
Rapid cavity ringdown measurements of multiple broadband absorbing species (methanol and isopropanol) in gas mixtures have been recorded with two multiplexed continuous-wave distributed-feedback diode lasers operating near 1.4 mum. A measurement sensitivity of 2.4 x 10(-9) cm(-1) for a 4.3-s averaging time was achieved in a 39.5-cm-long static cell with 99.94% reflectivity mirrors. This corresponds to a water-vapor detection limit of less than 2 ppb (parts in 10(9)) for the strong H(2)O lines near 1.4 mum. The shot-to-shot noise of the decay time constant tau was approximately 0.3-0.7%, and ringdown acquisition rates as great as 900 Hz were achieved.