» Articles » PMID: 18251617

Insertion Torque of Orthodontic Miniscrews According to Changes in Shape, Diameter and Length

Overview
Journal Angle Orthod
Specialty Dentistry
Date 2008 Feb 7
PMID 18251617
Citations 50
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To determine the variation in the insertion torque of orthodontic miniscrews according to the screw length, diameter, and shape.

Materials And Methods: The maximum insertion torque (MIT) was measured using a torque tester at a constant speed of 3 rotations per minute. Cylindrical and taper type of miniscrews (Biomaterials Korea Inc, Seoul, Korea) with different lengths, diameters, and pitches were tested.

Results: The results showed that the insertion torque significantly increased with increasing screw length (P < .01). In particular, there was a significant increase in torque with increasing screw length and diameter (P < .01). An analysis of the serial insertion torque of miniscrews revealed the cylindrical type screw to have much higher insertion torque at the incomplete screw thread, while the taper type screw showed a much higher insertion torque at the final inclination part of the screw thread. The insertion torque was affected by the outer diameter, length, and shape in that order.

Conclusions: An increase in screw diameter can efficiently reinforce the initial stability of the miniscrew, but the proximity of the root at the implanted site should be considered.

Citing Articles

Study on the Reasonability of Single-Objective Optimization in Miniscrew Design.

Li Y, Hwang J, Fung C Materials (Basel). 2025; 18(5).

PMID: 40077199 PMC: 11901241. DOI: 10.3390/ma18050973.


Anchorage and Stability of Orthodontic Mini Implants in Relation to Length and Types of Implants.

Selvaraj S, Tandon A, Chandrasekaran D, Purushothaman D, Katepogu P, Mohan R Cureus. 2024; 16(11):e73056.

PMID: 39640137 PMC: 11620767. DOI: 10.7759/cureus.73056.


Development and in vitro testing of an orthodontic miniscrew for use in the mandible.

Bauer C, Karl P, Mielke J, Roser C, Lux C, Scheurer M J Orofac Orthop. 2024; .

PMID: 39589499 DOI: 10.1007/s00056-024-00560-z.


Comparative assessment of the stability of buccal shelf mini-screws with and without pre-drilling- a split-mouth, randomized controlled trial.

Ummat A, Shetty S, Desai A, Nambiar S, Natarajan S Clin Oral Investig. 2024; 28(10):567.

PMID: 39365318 PMC: 11452459. DOI: 10.1007/s00784-024-05925-7.


Success rates of single-thread and double-thread orthodontic miniscrews in the maxillary arch.

Merati M, Ghaffari H, Javid F, Ahrari F BMC Oral Health. 2024; 24(1):191.

PMID: 38317101 PMC: 10845463. DOI: 10.1186/s12903-024-03866-x.