Melanometry for Objective Evaluation of Skin Pigmentation in Pulse Oximetry Studies

Overview
Publisher Nature Portfolio
Specialty General Medicine
Date 2024 Jul 11
PMID 38992188
Authors
Affiliations
Soon will be listed here.
Abstract

Pulse oximetry enables real-time, noninvasive monitoring of arterial blood oxygen levels. However, results can vary with skin color, thus detecting disparities during clinical validation studies requires an accurate measure of skin pigmentation. Recent clinical studies have used subjective methods such as self-reported color, race/ethnicity to categorize skin. Melanometers based on optical reflectance may offer a more effective, objective approach to assess pigmentation. Here, we review melanometry approaches and assess evidence supporting their use as clinical research tools. We compare performance data, including repeatability, robustness to confounders, and compare devices to each other, to subjective methods, and high-quality references. Finally, we propose best practices for evaluating melanometers and discuss alternate optical approaches that may improve accuracy. Whilst evidence indicates that melanometers can provide superior performance to subjective approaches, we encourage additional research and standardization efforts, as these are needed to ensure consistent and reliable results in clinical studies.

Citing Articles

Challenging the skin pigmentation bias in tissue oximetry via time-domain near-infrared spectroscopy.

Lacerenza M, Amendola C, Bargigia I, Bossi A, Buttafava M, Calcaterra V Biomed Opt Express. 2025; 16(2):690-708.

PMID: 39958842 PMC: 11828448. DOI: 10.1364/BOE.541239.


Adherence to FDA Guidance on Pulse Oximetry Testing Among Diverse Individuals, 1996-2024.

Ferryman K, Crews D, Drabo E, Iwashyna T, Kane O, Jackson J JAMA. 2025; 333(7):631-632.

PMID: 39786431 PMC: 11836753. DOI: 10.1001/jama.2024.26473.


Ninety years of pulse oximetry: history, current status, and outlook.

Quaresima V, Ferrari M, Scholkmann F J Biomed Opt. 2024; 29(Suppl 3):S33307.

PMID: 39156662 PMC: 11330276. DOI: 10.1117/1.JBO.29.S3.S33307.

References
1.
Vesoulis Z, Tims A, Lodhi H, Lalos N, Whitehead H . Racial discrepancy in pulse oximeter accuracy in preterm infants. J Perinatol. 2021; 42(1):79-85. PMC: 8508473. DOI: 10.1038/s41372-021-01230-3. View

2.
Osto M, Hamzavi I, Lim H, Kohli I . Individual Typology Angle and Fitzpatrick Skin Phototypes are Not Equivalent in Photodermatology. Photochem Photobiol. 2021; 98(1):127-129. DOI: 10.1111/php.13562. View

3.
STEWART K, Rowbottom S . Inaccuracy of pulse oximetry in patients with severe tricuspid regurgitation. Anaesthesia. 1991; 46(8):668-70. DOI: 10.1111/j.1365-2044.1991.tb09720.x. View

4.
Swiatoniowski A, Quillen E, Shriver M, Jablonski N . Technical note: comparing von Luschan skin color tiles and modern spectrophotometry for measuring human skin pigmentation. Am J Phys Anthropol. 2013; 151(2):325-30. DOI: 10.1002/ajpa.22274. View

5.
Saager R, Balu M, Crosignani V, Sharif A, Durkin A, Kelly K . In vivo measurements of cutaneous melanin across spatial scales: using multiphoton microscopy and spatial frequency domain spectroscopy. J Biomed Opt. 2015; 20(6):066005. PMC: 4463032. DOI: 10.1117/1.JBO.20.6.066005. View