Objective: To develop and determine the reliability of a newly -designed resistance-enhanced dynamometer for muscle strength measurement, and to test the hypothesis that enhancing the examiner's resisting force improves the reliability of manual muscle strength measurements.

Design: An intra-examiner, inter-examiner, intra-session and inter-session reliability study.

Subjects: Twenty-five men (mean age 22.5 (standard deviation (SD) 1.7) years) were tested separately by 2 examiners using the resistance-enhanced dynamometer and a traditional hand-held dynamometer for an intra- and inter-examiner reliability study. Twenty-seven volunteers (mean age 22.1 (SD 0.8) years) were tested by a female examiner using the resistance-enhanced dynamometer for an intra- and inter-session reliability study.

Methods: Maximum resisting forces for the knee flexors and extensors were measured using the resistance-enhanced dynamometer and the traditional hand-held dynamometer.

Results: The traditional hand-held dynamometer had good intra-examiner reliability (intra-class correlation coefficient (ICC) = 0.79-0.93) but poor inter-examiner reliability (ICC = 0.11-0.28). The resistance-enhanced dynamo meter had very good intra-examiner (ICC = 0.91-0.94), inter-examiner (ICC = 0.98), intra-session (ICC = 0.93-0.99) and inter-session (ICC = 0.91-0.92) reliability. The resistance-enhanced dynamometer also had better inter-examiner agreement (smallest real difference (SRD) 9-16% for resistance-enhanced dynamometer, 21-43% for traditional hand-held dynamometer).

Conclusion: The resistance-enhanced dynamometer had very good reliability. Enhancing the examiner's resisting force appeared to improve the reliability of manual muscle strength measurements. The resistance-enhanced dynamometer is useful for muscle strength measurements in clinical practice.

Citing Articles

Test-retest reliability and concurrent validity of knee extensor strength measured by a novel device incorporated into a weight stack machine vs. handheld and isokinetic dynamometry.

Sahu P, Goodstadt N, Ramakrishnan A, Silfies S PLoS One. 2024; 19(5):e0301872.

PMID: 38776288 PMC: 11111025. DOI: 10.1371/journal.pone.0301872.


Low-load blood flow restriction strength training in patients with COPD: a randomised single-blind pilot study.

Kohlbrenner D, Kuhn M, Manettas A, Aregger C, Peterer M, Greco N Thorax. 2023; 79(4):340-348.

PMID: 38129116 PMC: 10958309. DOI: 10.1136/thorax-2023-220546.


The effects of a two-week neuromuscular intervention on biopsychosocial variables in people with patellofemoral pain: an observational study.

Lack S, Bartholomew C, North T, Miller S, Neal B Front Sports Act Living. 2023; 5:1087061.

PMID: 37255732 PMC: 10225691. DOI: 10.3389/fspor.2023.1087061.


Interrater and Intrarater Reliability of the EasyForce Dynamometer for Assessment of Maximal Shoulder, Knee and Hip Strength.

Trajkovic N, Kozinc Z, Smajla D, Sarabon N Diagnostics (Basel). 2022; 12(2).

PMID: 35204532 PMC: 8871015. DOI: 10.3390/diagnostics12020442.


A prospective randomised control trial comparing functional with mechanical axis alignment in total knee arthroplasty: study protocol for an investigator initiated trial.

Steer R, Tippett B, Khan R, Collopy D, Clark G Trials. 2021; 22(1):523.

PMID: 34372888 PMC: 8351154. DOI: 10.1186/s13063-021-05433-z.