Dosimetric Impact of Tantalum Markers Used in the Treatment of Uveal Melanoma with Proton Beam Therapy

Overview
Journal Phys Med Biol
Publisher IOP Publishing
Date 2007 Aug 1
PMID 17664589
Citations 22
Authors
Affiliations
Soon will be listed here.
Abstract

Metallic fiducial markers are frequently implanted in patients prior to external-beam radiation therapy to facilitate tumor localization. There is little information in the literature, however, about the perturbations in proton absorbed-dose distribution these objects cause. The aim of this study was to assess the dosimetric impact of perturbations caused by 2.5 mm diameter by 0.2 mm thick tantalum fiducial markers when used in proton therapy for treating uveal melanoma. Absorbed dose perturbations were measured using radiochromic film and confirmed by Monte Carlo simulations of the experiment. Additional Monte Carlo simulations were performed to study the effects of range modulation and fiducial placement location on the magnitude of the dose shadow for a representative uveal melanoma treatment. The simulations revealed that the fiducials caused perturbations in the absorbed-dose distribution, including absorbed-dose shadows of 22% to 82% in a typical proton beam for treating uveal melanoma, depending on the marker depth and orientation. The clinical implication of this study is that implanted fiducials may, in certain circumstances, cause dose shadows that could lower the tumor dose and theoretically compromise local tumor control. To avoid this situation, fiducials should be positioned laterally or distally with respect to the target volume.

Citing Articles

Use of Carbon Fiber Implants to Improve the Safety and Efficacy of Radiation Therapy for Spine Tumor Patients.

Lam F, Guru S, AbuReesh D, Hori Y, Chuang C, Liu L Brain Sci. 2025; 15(2).

PMID: 40002531 PMC: 11852773. DOI: 10.3390/brainsci15020199.


Silicone Fiducial Markers Improve Precision in Uveal Melanoma Radiation Therapy.

Sonntag S, Wittenstein O, Blanck O, Dunst J, Huttenlocher S, Grehn M Cancers (Basel). 2025; 17(2).

PMID: 39857971 PMC: 11763580. DOI: 10.3390/cancers17020189.


A Novel Polymer-Encapsulated Multi-Imaging Modality Fiducial Marker with Positive Signal Contrast for Image-Guided Radiation Therapy.

Wang L, Sanders J, Ward J, Lee S, Poenisch F, Swanson D Cancers (Basel). 2024; 16(3).

PMID: 38339376 PMC: 10854757. DOI: 10.3390/cancers16030625.


Identification of Induced Radionuclides Produced from Dental Metals in Proton Beam Therapy for Head and Neck Cancer.

Kato R, Kato T, Narita Y, Sasaki S, Takayama K, Murakami M Adv Radiat Oncol. 2023; 8(4):101153.

PMID: 36798730 PMC: 9926195. DOI: 10.1016/j.adro.2022.101153.


Experimental Comparison of Fiducial Markers Used in Proton Therapy: Study of Different Imaging Modalities and Proton Fluence Perturbations Measured With CMOS Pixel Sensors.

Reidel C, Horst F, Schuy C, Jakel O, Ecker S, Henkner K Front Oncol. 2022; 12:830080.

PMID: 35402273 PMC: 8990863. DOI: 10.3389/fonc.2022.830080.