Accuracy of Blood Flow Values Determined by Arterial Spin Labeling: a Validation Study in Isolated Porcine Kidneys

Overview
Date 2007 Jul 27
PMID 17654732
Citations 12
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To validate the accuracy of quantitative blood flow values determined using pulsed arterial spin labeling (ASL) in the preserved and reperfused porcine kidney.

Materials And Methods: Ex vivo porcine kidneys were perfused with whole blood under physiological conditions, in particular including pulsatile flow. Total flow through the kidney was determined using an ultrasound flowmeter. ASL measurements at two different inversion times and four different flow rates in the range of 70-210 mL/100 mL*minute were performed. Absolute values of blood flow and arterial transit times were determined in the kidney cortex.

Results: The quantitative values were in good agreement with the reference values obtained after calibration of the total flow. The greatest difference observed was 13%.

Conclusion: Isolated organ hemoperfusion allows validating perfusion imaging techniques. The experimental setup enables long-term radiotherapeutic or toxicological studies using noninvasive ASL to monitor blood flow quantitatively.

Citing Articles

Assessment of pharmacologically induced changes in canine kidney function by multiparametric magnetic resonance imaging and contrast enhanced ultrasound.

Hillaert A, Sanmiguel Serpa L, Bogaert S, Broeckx B, Hesta M, Vandermeulen E Front Vet Sci. 2024; 11:1406343.

PMID: 38966564 PMC: 11223176. DOI: 10.3389/fvets.2024.1406343.


Arterial Spin Labeling Magnetic Resonance Imaging for Acute Disorders of Consciousness in the Intensive Care Unit.

Gronlund E, Lindberg U, Fisher P, Othman M, Amiri M, Solling C Neurocrit Care. 2024; 41(3):1027-1037.

PMID: 38918338 PMC: 11599417. DOI: 10.1007/s12028-024-02031-0.


Three-dimensional US Fractional Moving Blood Volume: Validation of Renal Perfusion Quantification.

Welsh A, Fowlkes J, Pinter S, Ives K, Owens G, Rubin J Radiology. 2019; 293(2):460-468.

PMID: 31573404 PMC: 6800606. DOI: 10.1148/radiol.2019190248.


Arterial spin labelling MRI to measure renal perfusion: a systematic review and statement paper.

Odudu A, Nery F, Harteveld A, Evans R, Pendse D, Buchanan C Nephrol Dial Transplant. 2018; 33(suppl_2):ii15-ii21.

PMID: 30137581 PMC: 6106644. DOI: 10.1093/ndt/gfy180.


Non-Invasive Renal Perfusion Imaging Using Arterial Spin Labeling MRI: Challenges and Opportunities.

Nery F, Gordon I, Thomas D Diagnostics (Basel). 2018; 8(1).

PMID: 29303965 PMC: 5871985. DOI: 10.3390/diagnostics8010002.