An Assessment of the Use of Skin Flashes in Helical Tomotherapy Using Phantom and In-vivo Dosimetry

Overview
Journal Radiother Oncol
Specialties Oncology
Radiology
Date 2007 Jun 26
PMID 17590468
Citations 11
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Purpose: In helical tomotherapy the nature of the optimizing and planning systems allows the delivery of dose on the skin using a build-up compensating technique (skin flash). However, positioning errors or changes in the patient's contour can influence the correct dosage in these regions. This work studies the behavior of skin-flash regions using phantom and in-vivo dosimetry.

Materials And Methods: The dosimetric accuracy of the tomotherapy planning system in skin-flash regions is checked using film and TLD on phantom. Positioning errors are induced and the effect on the skin dose is investigated. Further a volume decrease is simulated using bolus material and the results are compared.

Results: Results show that the tomotherapy planning system calculates dose on skin regions within 2 SD using TLD measurements. Film measurements show drops of dose of 2.8% and 26% for, respectively, a 5mm and 10mm mispositioning of the phantom towards air and a dose increase of 9% for a 5mm shift towards tissue. These measurements are confirmed by TLD measurements. A simulated volume reduction shows a similar behavior with a 2.6% and 19.4% drop in dose, measured with TLDs.

Conclusion: The tomotherapy system allows adequate planning and delivery of dose using skin flashes. However, exact positioning is crucial to deliver the dose at the exact location.

Citing Articles

A dosiomics model for prediction of radiation-induced acute skin toxicity in breast cancer patients: machine learning-based study for a closed bore linac.

Saadatmand P, Mahdavi S, Nikoofar A, Jazaeri S, Ramandi F, Esmaili G Eur J Med Res. 2024; 29(1):282.

PMID: 38735974 PMC: 11089719. DOI: 10.1186/s40001-024-01855-y.


The effects of rotational setup errors in total body irradiation using helical tomotherapy.

Isobe A, Usui K, Hara N, Sasai K J Appl Clin Med Phys. 2021; 22(7):93-102.

PMID: 34028944 PMC: 8292714. DOI: 10.1002/acm2.13271.


Skin dose calculation during radiotherapy of head and neck cancer using deformable image registration of planning and mega-voltage computed tomography scans.

Branchini M, Broggi S, DellOca I, Cattaneo G, Calandrino R, Di Muzio N Phys Imaging Radiat Oncol. 2021; 8:44-50.

PMID: 33458416 PMC: 7807680. DOI: 10.1016/j.phro.2018.11.008.


Clinical implementation of low-dose total body irradiation using topotherapy technique.

Broggi S, Fiorino C, Chiara A, Salvadori G, Peccatori J, Assanelli A Phys Imaging Radiat Oncol. 2021; 12:74-79.

PMID: 33458299 PMC: 7807637. DOI: 10.1016/j.phro.2019.11.009.


Dosimetric effects of anatomical deformations and positioning errors in VMAT breast radiotherapy.

Rossi M, Boman E, Skytta T, Haltamo M, Laaksomaa M, Kapanen M J Appl Clin Med Phys. 2018; 19(5):506-516.

PMID: 29978548 PMC: 6123165. DOI: 10.1002/acm2.12409.