A New Device for Measuring Resting Energy Expenditure (REE) in Healthy Subjects

Overview
Publisher Elsevier
Date 2007 Jun 15
PMID 17562571
Citations 34
Authors
Affiliations
Soon will be listed here.
Abstract

Background And Aim: Lifestyle change targeted towards increasing daily resting energy expenditure (REE) is one of the cornerstones of obesity treatment. Measurements of energy expenditure and substrate utilization are essential to understanding the metabolic basis of obesity, and the physiological responses to perturbations in habitual food intake. REE is the largest part of human energy expenditure (60-70%) and an increase or decrease in REE would have a large impact on total energy. Accurate and easy-to-use methods for measuring REE are needed, to be applied by clinicians in daily clinical settings to assess the validity of a new instrument to estimate REE in normal weight, healthy adults.

Methods: Ninety-nine subjects (52 females and 47 males) (mean+/-SD, age 38+/-14 years; body mass index (BMI) 23+/-3 kg/m(2)) were tested. REE was assessed using a Sensor Medics Vmax metabolic cart with a ventilated canopy and with the SenseWear armband. Body composition, percentage fat mass (%FM) and percentage fat free mass (%FFM) were assessed by skinfold thickness measurements (SF), bio-electrical impedance analysis (BIA) and air displacement plethysmography (BOD-POD).

Results: No significant difference was found among measurements of FFM using the three different techniques. Both SenseWear and Sensor Medics Vmax showed a high correlation, r=0.42 and r=0.40 (p<0.0001) respectively, with BMI. No significant difference was found in mean REE between SenseWear (1540+/-280 kcal/day) and Sensor Medics Vmax (1700+/-330 kcal/day) (p=ns) and the correlation between REE measured by SenseWear and Sensor Medics Vmax was high (r=0.86, p<0.0001). Bland-Altman plot showed no difference in REE determination between SenseWear and Sensor Medics Vmax. %FFM determined by BOD-POD correlated with SenseWear (r=0.42, p<0.0001) as well as Sensor Medics Vmax (r=0.38, p<0.001).

Conclusion: SF, BIA and BOD-POD provide valid and reliable measurements of FFM. Our results suggest that the SenseWear armband is an acceptable device to accurately measure REE in healthy subjects. Its characteristics have the potential to reduce measurement times and make the SenseWear armband useful for epidemiological studies.

Citing Articles

Augmented Rehabilitation Program for Patients 60 Years and Younger Following Total Hip Arthroplasty-Feasibility Study.

Negm A, Yavarai M, Jhangri G, Haennel R, Jones C Healthcare (Basel). 2022; 10(7).

PMID: 35885801 PMC: 9324868. DOI: 10.3390/healthcare10071274.


Validity of Accelerometers for the Evaluation of Energy Expenditure in Obese and Overweight Individuals: A Systematic Review.

Pisanu S, Deledda A, Loviselli A, Huybrechts I, Velluzzi F J Nutr Metab. 2020; 2020:2327017.

PMID: 32832147 PMC: 7424495. DOI: 10.1155/2020/2327017.


Is There an Association or Not?-Investigating the Association of Depressiveness, Physical Activity, Body Composition and Sleep With Mediators of Inflammation.

Schmidt F, Mergl R, Minkwitz J, Holdt L, Teupser D, Hegerl U Front Psychiatry. 2020; 11:563.

PMID: 32670105 PMC: 7330708. DOI: 10.3389/fpsyt.2020.00563.


Low-Cost Consumer-Based Trackers to Measure Physical Activity and Sleep Duration Among Adults in Free-Living Conditions: Validation Study.

Degroote L, Hamerlinck G, Poels K, Maher C, Crombez G, De Bourdeaudhuij I JMIR Mhealth Uhealth. 2020; 8(5):e16674.

PMID: 32282332 PMC: 7268004. DOI: 10.2196/16674.


The effects of diurnal intermittent fasting on proinflammatory cytokine levels while controlling for sleep/wake pattern, meal composition and energy expenditure.

Almeneessier A, BaHammam A, Alzoghaibi M, Olaish A, Nashwan S, BaHammam A PLoS One. 2019; 14(12):e0226034.

PMID: 31821377 PMC: 6903761. DOI: 10.1371/journal.pone.0226034.