Oxygen Permeability and Water Content of Silicone Hydrogel Contact Lens Materials
Overview
Affiliations
Purpose: To measure the oxygen permeability (Dk) and water content (WC) of silicone hydrogel (Si-Hy) contact lens materials.
Methods: Randomized and masked determinations of the Dk of 5 Si-Hy and two hydrogel materials were made using a modified version of the polarographic measurement method described in ISO 9913-1. Stacks of one to six parallel-sided contact lenses (all -1.00 DS) were evaluated, with each stack measured at least twice. The resulting value for t/Dk was plotted against the thickness (t) of each stack, with Dk calculated as the inverse of the gradient of this relationship. This methodology corrects for boundary effects. A mathematical calculation was used to correct for edge effects. Gravimetric determination of lens WC was conducted at room temperature and 35 degrees C.
Results: Measured values (+/-95% CI) of Dk, and WC at room temperature, with manufacturer-claimed values in parentheses, were Focus Night and Day: Dk 162.0 +/- 9.8 (140), WC 23.0 +/- 3.2 (24); Acuvue Oasys: Dk 107.4 +/- 7.4 (103), WC 36.9 +/- 1.0 (38); O2 Optix: Dk 80.5 +/- 4.9 (110), WC 32.1 +/- 1.2 (33); PureVision: Dk 75.9 +/- 6.6 (91), WC 35.8 +/- 1.3 (36); Acuvue Advance: Dk 75.2 +/- 9.8 (60), WC 46.5 +/- 1.1 (47); 1. Day Acuvue: Dk 21.0 +/- 1.0 (21.4), WC [not measured] (58); and Seequence: Dk 8.2 +/- 0.7 (8.5), WC 36.6 +/- 2.7 (38).
Conclusions: Claimed Dk values for Acuvue Oasys and the two reference hydrogel materials fell within the 95% confidence interval of our measured values. Our measurements of Dk for the other four Si-Hy lenses were not in agreement with claimed values. There is a general inverse relation between Dk and WC (both at 35 degrees C) for Si-Hy lenses. Our modified polarographic methodology can be successfully employed for measuring the Dk of Si-Hy materials.
Biomimetic-Engineered Silicone Hydrogel Contact Lens Materials.
Ishihara K, Shi X, Fukazawa K, Yamaoka T, Yao G, Wu J ACS Appl Bio Mater. 2023; 6(9):3600-3616.
PMID: 37616500 PMC: 10521029. DOI: 10.1021/acsabm.3c00296.
The pH-induced physical properties of ionic contact lens material.
Ahn J, Choi M Heliyon. 2023; 9(2):e12996.
PMID: 36793947 PMC: 9922922. DOI: 10.1016/j.heliyon.2023.e12996.
Smart Wireless Near-Infrared Light Emitting Contact Lens for the Treatment of Diabetic Retinopathy.
Lee G, Jeon C, Mok J, Shin S, Kim S, Han H Adv Sci (Weinh). 2022; 9(9):e2103254.
PMID: 35092362 PMC: 8948592. DOI: 10.1002/advs.202103254.
Efron N, Jones L, Morgan P, Nichols J J Optom. 2021; 15(1):44-52.
PMID: 34615612 PMC: 8712540. DOI: 10.1016/j.optom.2021.05.003.
Zargarnezhad H, Asselin E, Wong D, Lam C Polymers (Basel). 2021; 13(9).
PMID: 34065062 PMC: 8125940. DOI: 10.3390/polym13091517.