Optimal Apodization Design for Medical Ultrasound Using Constrained Least Squares Part II: Simulation Results

Overview
Date 2007 Mar 3
PMID 17328331
Citations 10
Authors
Affiliations
Soon will be listed here.
Abstract

In the first part of this work, we introduced a novel general ultrasound apodization design method using constrained least squares (CLS). The technique allows for the design of system spatial impulse responses with narrow mainlobes and low sidelobes. In the linear constrained least squares (LCLS) formulation, the energy of the point spread function (PSF) outside a certain mainlobe boundary was minimized while maintaining a peak gain at the focus. In the quadratic constrained least squares (QCLS) formulation, the energy of the PSF outside a certain boundary was minimized, and the energy of the PSF inside the boundary was held constant. In this paper, we present simulation results that demonstrate the application of the CLS methods to obtain optimal system responses. We investigate the stability of the CLS apodization design methods with respect to errors in the assumed wave propagation speed. We also present simulation results that implement the CLS design techniques to improve cystic resolution. According to novel performance metrics, our apodization profiles improve cystic resolution by 3 dB to 10 dB over conventional apodizations such as the flat, Hamming, and Nuttall windows. We also show results using the CLS techniques to improve conventional depth of field (DOF).

Citing Articles

Grating Lobe Reduction in Plane-Wave Imaging With Angular Compounding Using Subtraction of Coherent Signals.

Kou Z, Miller R, Oelze M IEEE Trans Ultrason Ferroelectr Freq Control. 2022; 69(12):3308-3316.

PMID: 36306299 PMC: 9741941. DOI: 10.1109/TUFFC.2022.3217993.


Improving non-line-of-sight image reconstruction with weighting factors.

Feng X, Gao L Opt Lett. 2020; 45(14):3921-3924.

PMID: 32667319 PMC: 7416968. DOI: 10.1364/OL.394742.


Improving Spatial Resolution Using Incoherent Subtraction of Receive Beams Having Different Apodizations.

Agarwal A, Reeg J, Podkowa A, Oelze M IEEE Trans Ultrason Ferroelectr Freq Control. 2018; 66(1):5-17.

PMID: 30334791 PMC: 6350248. DOI: 10.1109/TUFFC.2018.2876285.


A Robust Method for Ultrasound Beamforming in the Presence of Off-Axis Clutter and Sound Speed Variation.

Dei K, Byram B Ultrasonics. 2018; 89:34-45.

PMID: 29723842 PMC: 6014921. DOI: 10.1016/j.ultras.2018.04.011.


A sidelobe suppressing near-field beamforming approach for ultrasound array imaging.

He Z, Zheng F, Ma Y, Kim H, Zhou Q, Shung K J Acoust Soc Am. 2015; 137(5):2785-90.

PMID: 25994706 PMC: 5392098. DOI: 10.1121/1.4919318.