Estimation of Effective Doses to Adult and Pediatric Patients from Multislice Computed Tomography: A Method Based on Energy Imparted

Overview
Journal Med Phys
Specialty Biophysics
Date 2006 Nov 9
PMID 17089849
Citations 13
Authors
Affiliations
Soon will be listed here.
Abstract

The purpose of this study is to provide a method and required data for the estimation of effective dose (E) values to adult and pediatric patients from computed tomography (CT) scans of the head, chest abdomen, and pelvis, performed on multi-slice scanners. Mean section radiation dose (dm) to cylindrical water phantoms of varying radius normalized over CT dose index free-in-air (CTDIF) were calculated for the head and body scanning modes of a multislice scanner with use of Monte Carlo techniques. Patients were modeled as equivalent water phantoms and the energy imparted (epsilon) to simulated pediatric and adult patients was calculated on the basis of measured CTDI(F) values. Body region specific energy imparted to effective dose conversion coefficients (E/epsilon) for adult male and female patients were generated from previous data. Effective doses to patients aged newborn to adult were derived for all available helical and axial beam collimations, taking into account age specific patient mass and scanning length. Depending on high voltage, body region, and patient sex, E/epsilon values ranged from 0.008 mSv/mJ for head scans to 0.024 mSv/mJ for chest scans. When scanned with the same technique factors as the adults, pediatric patients absorb as little as 5% of the energy imparted to adults, but corresponding effective dose values are up to a factor of 1.6 higher. On average, pediatric patients absorb 44% less energy per examination but have a 24% higher effective dose, compared with adults. In clinical practice, effective dose values to pediatric patients are 2.5 to 10 times lower than in adults due to the adaptation of tube current. A method is provided for the calculation of effective dose to adult and pediatric patients on the basis of individual patient characteristics such as sex, mass, dimensions, and density of imaged anatomy, and the technical features of modern multislice scanners. It allows the optimum selection of scanning parameters regarding patient doses at CT.

Citing Articles

Introducing fitting models for estimating age-specific dose and effective dose in paediatric patients undergoing head, chest and abdomen-pelvis imaging protocols: a patient study.

Deevband M, Mohammadi H, Salimi Y, Mostaar A, Deravi N, Fathi M J Med Radiat Sci. 2024; 71(2):251-260.

PMID: 38454637 PMC: 11177019. DOI: 10.1002/jmrs.772.


Radiographic and fluoroscopic X-ray systems: Quality control of the X-ray tube and automatic exposure control using theoretical spectra to determine air kerma and dose to a homogenous phantom.

Konst B, Notthellen J, Bilet E, Bath M J Appl Clin Med Phys. 2021; 22(8):204-218.

PMID: 34196461 PMC: 8364276. DOI: 10.1002/acm2.13329.


Radiation dosimetry of 18F-FDG PET/CT: incorporating exam-specific parameters in dose estimates.

Quinn B, Dauer Z, Pandit-Taskar N, Schoder H, Dauer L BMC Med Imaging. 2016; 16(1):41.

PMID: 27317478 PMC: 4912712. DOI: 10.1186/s12880-016-0143-y.


Evaluation of diagnostic performance of whole-body simultaneous PET/MRI in pediatric lymphoma.

Ponisio M, McConathy J, Laforest R, Khanna G Pediatr Radiol. 2016; 46(9):1258-68.

PMID: 27003132 PMC: 5841580. DOI: 10.1007/s00247-016-3601-3.


Value of diffusion-weighted MR imaging for the detection of nephritis.

Henninger B, Reichert M, Haneder S, Schoenberg S, Michaely H ScientificWorldJournal. 2013; 2013:348105.

PMID: 24319359 PMC: 3844243. DOI: 10.1155/2013/348105.