Reproducibility of Target Volumes Generated Using Uncoached 4-dimensional CT Scans for Peripheral Lung Cancer

Overview
Journal Radiat Oncol
Publisher Biomed Central
Specialties Oncology
Radiology
Date 2006 Nov 3
PMID 17078882
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

Background: 4-dimensional CT (4DCT) scans are increasingly used to account for mobility during radiotherapy planning. As variations in respiratory patterns can alter observed motion, with consequent changes in the generated target volumes, we evaluated the reproducibility of 4D target volumes generated during repeat uncoached quiet respiration.

Methods: A retrospective analysis was performed on two successive scans (4DCT1 and 4DCT2) generated at the same scanning session for 26 patients with peripheral lung cancer treated with stereotactic radiotherapy (SRT). The volume and position of planning target volumes (PTV4DCT1 and PTV4DCT2) contoured on both scans were compared, and a dosimetric analysis performed. A SRT plan optimized for each PTV was sequentially applied to the other PTV, and coverage by the 80% isodose was evaluated. Color intensity projections (CIP) were used to evaluate regions of underdosage.

Results: No significant volumetric differences were observed between the two PTVs (t-Test p = 0.60). The average displacement of the center of mass between corresponding PTVs was 1.4 +/- 1.0 mm, but differences in position were 2.0 mm or greater in 5 cases (19%). Coverage of both PTVs by the 80% prescription isodose exceeded 90% for all but one patient. For the latter, the prescription isodose covered only 82.5% of PTV4DCT1. CIP analysis revealed that the region of underdosage was an end-inspiratory position occupied by the tumor for only 10-20% of the respiratory cycle.

Conclusion: In nearly all patients with stage I lung cancer, the PTV derived from a single uncoached 4DCT achieves dosimetric coverage that is similar to that achieved using two such consecutive scans.

Citing Articles

Lung stereotactic body radiation therapy: personalized PTV margins according to tumor location and number of four-dimensional CT scans.

Tremolieres P, Gonzalez-Moya A, Paumier A, Mege M, Blanchecotte J, Theotime C Radiat Oncol. 2022; 17(1):5.

PMID: 35012579 PMC: 8751327. DOI: 10.1186/s13014-021-01973-5.


Technical Considerations in Stereotactic Ablative Radiotherapy for Localized Neuroendocrine Cancer of the Lung: Case Report and Review of the Literature.

Callan L, Gaede S, Louie A Adv Radiat Oncol. 2019; 4(1):26-30.

PMID: 30706006 PMC: 6349633. DOI: 10.1016/j.adro.2018.10.002.


The impact of anatomic tumor location on inter-fraction tumor motion during lung stereotactic body radiation therapy (SBRT).

Atkins K, Chen Y, Elliott D, Doshi T, Ognjenovic S, Vachhani A J Radiosurg SBRT. 2018; 3(3):203-213.

PMID: 29296403 PMC: 5746335.


Respiratory gating during stereotactic body radiotherapy for lung cancer reduces tumor position variability.

Saito T, Matsuyama T, Toya R, Fukugawa Y, Toyofuku T, Semba A PLoS One. 2014; 9(11):e112824.

PMID: 25379729 PMC: 4224502. DOI: 10.1371/journal.pone.0112824.


Sequential dual-energy subtraction technique with a dynamic flat-panel detector (FPD): primary study for image-guided radiation therapy (IGRT).

Tanaka R, Sanada S, Matsui T, Hayashi N, Matsui O Radiol Phys Technol. 2010; 1(2):144-50.

PMID: 20821140 DOI: 10.1007/s12194-008-0021-6.


References
1.
Van Herk M, Witte M, van der Geer J, Schneider C, Lebesque J . Biologic and physical fractionation effects of random geometric errors. Int J Radiat Oncol Biol Phys. 2003; 57(5):1460-71. DOI: 10.1016/j.ijrobp.2003.08.026. View

2.
van Sornsen de Koste J, Lagerwaard F, Nijssen-Visser M, Graveland W, Senan S . Tumor location cannot predict the mobility of lung tumors: a 3D analysis of data generated from multiple CT scans. Int J Radiat Oncol Biol Phys. 2003; 56(2):348-54. DOI: 10.1016/s0360-3016(02)04467-x. View

3.
Keall P . 4-dimensional computed tomography imaging and treatment planning. Semin Radiat Oncol. 2004; 14(1):81-90. DOI: 10.1053/j.semradonc.2003.10.006. View

4.
Pan T, Lee T, Rietzel E, Chen G . 4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT. Med Phys. 2004; 31(2):333-40. DOI: 10.1118/1.1639993. View

5.
Keall P, Starkschall G, Shukla H, Forster K, Ortiz V, Stevens C . Acquiring 4D thoracic CT scans using a multislice helical method. Phys Med Biol. 2004; 49(10):2053-67. DOI: 10.1088/0031-9155/49/10/015. View