Four-dimensional CT Scans for Treatment Planning in Stereotactic Radiotherapy for Stage I Lung Cancer

Overview
Specialties Oncology
Radiology
Date 2004 Nov 3
PMID 15519801
Citations 63
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: Hypofractionated stereotactic radiotherapy (SRT) for Stage I non-small-cell lung cancer requires that meticulous attention be paid toward ensuring optimal target definition. Two computed tomography (CT) scan techniques for defining internal target volumes (ITV) were evaluated.

Methods And Materials: Ten consecutive patients treated with SRT underwent six "standard" rapid multislice CT scans to generate an ITV(6 CT) and one four-dimensional CT (4DCT) scan that generated volumetric datasets for 10 phases of the respiratory cycle, all of which were used to generate an ITV(4DCT). Geometric and dosimetric analyses were performed for (1) PTV(4DCT), derived from the ITV(4DCT) with the addition of a 3-mm margin; (2) PTV(6 CT), derived from the ITV(6 CT) with the addition of a 3-mm margin; and (3) 6 PTV(10 mm), derived from each separate GTV(6 CT), to which a three-dimensional margin of 10 mm was added.

Results: The ITV(4DCT) was not significantly different from the ITV(6 CT) in 8 patients, but was considerably larger in 2 patients whose tumors exhibited the greatest mobility. On average, the ITV(6 CT) missed on average 22% of the volume encompassing both ITVs, in contrast to a corresponding mean value of only 8.3% for ITV(4DCT). Plans based on PTV(4DCT) resulted in coverage of the PTV(6 CT) by the 80% isodose in all patients. However, plans based on use of PTV(6 CT) led to a mean PTV(4DCT) coverage of only 92.5%, with a minimum of 77.7% and 77.5% for the two most mobile tumors. PTVs derived from a single multislice CT expanded with a margin of 10 mm were on average twice the size of PTVs derived using the other methods, but still led to an underdosing in the two most mobile tumors.

Conclusions: Individualized ITVs can improve target definition for SRT of Stage I non-small-cell lung cancer, and use of only a single CT scan with a 10-mm margin is inappropriate. A single 4D scan generates comparable or larger ITVs than are generated using six unmonitored rapid CT scans, a finding related to the ability to account for all respiration-correlated mobility.

Citing Articles

Dosimetric comparison of four-dimensional computed tomography based internal target volume against variations in respiratory motion during treatment between volumetric modulated arc therapy and three-dimensional conformal radiotherapy in lung....

Fujimoto D, Takatsu J, Hara N, Oshima M, Tomihara J, Segawa E Radiol Phys Technol. 2023; 17(1):143-152.

PMID: 37930563 DOI: 10.1007/s12194-023-00757-8.


Investigation of uncertainty in internal target volume definition for lung stereotactic body radiotherapy.

Nakanishi D, Oita M, Fukunaga J, Hirose T, Yoshitake T, Sasaki M Radiol Phys Technol. 2023; 16(4):497-505.

PMID: 37713060 PMC: 10665452. DOI: 10.1007/s12194-023-00737-y.


Endobronchially Implanted Real-Time Electromagnetic Transponder Beacon-Guided, Respiratory-Gated SABR for Moving Lung Tumors: A Prospective Phase 1/2 Cohort Study.

Cheng J, Buduhan G, Venkataraman S, Tan L, Sasaki D, Bashir B Adv Radiat Oncol. 2023; 8(5):101243.

PMID: 37408673 PMC: 10318214. DOI: 10.1016/j.adro.2023.101243.


Film measurement and analytical approach for assessing treatment accuracy and latency in a magnetic resonance-guided radiotherapy system.

Nakayama H, Okamoto H, Nakamura S, Iijima K, Chiba T, Takemori M J Appl Clin Med Phys. 2023; 24(5):e13915.

PMID: 36934441 PMC: 10161048. DOI: 10.1002/acm2.13915.


Adaptive hypofractionted and stereotactic body radiotherapy for lung tumors with real-time MRI guidance.

Bryant J, Sim A, Feygelman V, Latifi K, Rosenberg S Front Oncol. 2023; 13:1061854.

PMID: 36776319 PMC: 9911810. DOI: 10.3389/fonc.2023.1061854.