Background: A new eight-wavelength pulse oximeter is designed to measure methemoglobin and carboxyhemoglobin, in addition to the usual measurements of hemoglobin oxygen saturation and pulse rate. This study examines this device's ability to measure dyshemoglobins in human volunteers in whom controlled levels of methemoglobin and carboxyhemoglobin are induced.

Methods: Ten volunteers breathed 500 ppm carbon monoxide until their carboxyhemoglobin levels reached 15%, and 10 different volunteers received intravenous sodium nitrite, 300 mg, to induce methemoglobin. All were instrumented with arterial cannulas and six Masimo Rad-57 (Masimo Inc., Irvine, CA) pulse oximeter sensors. Arterial blood was analyzed by three laboratory CO-oximeters, and the resulting carboxyhemoglobin and methemoglobin measurements were compared with the corresponding pulse oximeter readings.

Results: The Rad-57 measured carboxyhemoglobin with an uncertainty of +/-2% within the range of 0-15%, and it measured methemoglobin with an uncertainty of 0.5% within the range of 0-12%.

Conclusion: The Masimo Rad-57 is the first commercially available pulse oximeter that can measure methemoglobin and carboxyhemoglobin, and it therefore represents an expansion of our oxygenation monitoring capability.

Citing Articles

Environmental sustainability in the operating room: A worldwide survey among anaesthesiologists.

Gonzalez-Pizarro P, Koch S, Muret J, Trinks A, Brazzi L, Reinoso-Barbero F Eur J Anaesthesiol Intensive Care. 2025; 2(4):e0025.

PMID: 39917069 PMC: 11783648. DOI: 10.1097/EA9.0000000000000025.


Diagnosis of carbon monoxide exposure in clinical research and practice: A scoping review.

Moss P, Matthews N, McDonald R, Jarman H PLoS One. 2025; 20(2):e0300989.

PMID: 39908298 PMC: 11798492. DOI: 10.1371/journal.pone.0300989.


Correlation between Carboxyhemoglobin Levels Measured by Blood Gas Analysis and by Multiwave Pulse Oximetry.

Yu J, Lee J, Cho Y, Oh J, Kang H, Lim T J Pers Med. 2024; 14(2).

PMID: 38392600 PMC: 10890311. DOI: 10.3390/jpm14020168.


Evaluation of a Portable Blood Gas Analyzer for Prehospital Triage in Carbon Monoxide Poisoning: Instrument Validation Study.

Lyon M, Fehlmann C, Augsburger M, Schaller T, Zimmermann-Ivol C, Celi J JMIR Form Res. 2023; 7:e48057.

PMID: 37801355 PMC: 10589834. DOI: 10.2196/48057.


Exploring the Anti-Hypoxaemia Effect of Hydromethylthionine: A Prospective Study of Phase 3 Clinical Trial Participants.

Arastoo M, Mazanetz M, Miller S, Shiells H, Hull C, Robinson K Int J Mol Sci. 2023; 24(18).

PMID: 37762050 PMC: 10531415. DOI: 10.3390/ijms241813747.