The Long-term Stability of Portable Spirometers Used in a Multinational Study of the Prevalence of Chronic Obstructive Pulmonary Disease

Overview
Journal Respir Care
Publisher Mary Ann Liebert
Specialty Pulmonary Medicine
Date 2006 Sep 29
PMID 17005063
Citations 31
Authors
Affiliations
Soon will be listed here.
Abstract

Background: We report the performance of an ultrasound-based portable spirometer (EasyOne) used in a population-based survey of the prevalence of chronic obstructive pulmonary disease, conducted in 5 Latin American cities: São Paulo, Brazil; México City, México; Montevideo, Uruguay; Santiago, Chile; and Caracas, Venezuela (the Latin American COPD Prevalence Study [PLATINO]).

Methods: During the survey period (which ranged from 3 months to 6 months in the various locations) we collected daily calibration data from the 70 EasyOne spirometers used in the 5 survey cities. The calibrations were conducted with a 3-L syringe, and the calibration data were stored in the spirometer's database.

Results: Ninety-seven percent of the calibration volumes were within +/- 64 mL (2.1%) of the 3-L calibration signal. Excluding data from the first city studied (São Paulo), where one calibration syringe had to be replaced, 98% of the calibration checks were within +/- 50 mL (1.7%). The measured volume was affected only minimally by the syringe's peak flow or emptying time.

Conclusion: In these 70 EasyOne spirometers neither calibration nor linearity changed during the study. Such calibration stability is a valuable feature in spirometry surveys and in the clinical setting.

Citing Articles

Chronic obstructive Pulmonary Disease (COPD) in non-smoking Sri Lankan adults; a cross-sectional study.

Gunasekera K, Amarasiri W, Wickremasinghe A, Perera B, Undugodage U, Fernando A BMC Res Notes. 2025; 18(1):84.

PMID: 40001249 PMC: 11863925. DOI: 10.1186/s13104-025-07146-x.


Protocol for the air purification for eosinophilic COPD study (APECS): a randomised controlled trial of home air filtration by HEPA.

Saeed M, Denoncourt C, Chao I, Schortmann S, Nassikas N, Synn A BMJ Open. 2024; 14(1):e074655.

PMID: 38238060 PMC: 10806745. DOI: 10.1136/bmjopen-2023-074655.


Extracellular Vesicle-Encapsulated microRNAs as Novel Biomarkers of Lung Health.

Eckhardt C, Gambazza S, Bloomquist T, de Hoff P, Vuppala A, Vokonas P Am J Respir Crit Care Med. 2022; 207(1):50-59.

PMID: 35943330 PMC: 9952856. DOI: 10.1164/rccm.202109-2208OC.


Restricted spirometry and cardiometabolic comorbidities: results from the international population based BOLD study.

Kulbacka-Ortiz K, Triest F, Franssen F, Wouters E, Studnicka M, Vollmer W Respir Res. 2022; 23(1):34.

PMID: 35177082 PMC: 8855577. DOI: 10.1186/s12931-022-01939-5.


Increased pulmonary artery diameter is associated with reduced FEV in former World Trade Center workers.

de la Hoz R, Jeon Y, Reeves A, San Jose Estepar R, Liu X, Doucette J Clin Respir J. 2019; 13(10):614-623.

PMID: 31347281 PMC: 6783324. DOI: 10.1111/crj.13067.