Abstract

Introduction: With the 3100B high-frequency oscillatory ventilator (SensorMedics, Yorba Linda, CA, USA), patients' spontaneous breathing efforts result in a high level of imposed work of breathing (WOB). Therefore, spontaneous breathing often has to be suppressed during high-frequency oscillatory ventilation (HFOV). A demand-flow system was designed to reduce imposed WOB.

Methods: An external gas flow controller (demand-flow system) accommodates the ventilator fresh gas flow during spontaneous breathing simulation. A control algorithm detects breathing effort and regulates the demand-flow valve. The effectiveness of this system has been evaluated in a bench test. The Campbell diagram and pressure time product (PTP) are used to quantify the imposed workload.

Results: Using the demand-flow system, imposed WOB is considerably reduced. The demand-flow system reduces inspiratory imposed WOB by 30% to 56% and inspiratory imposed PTP by 38% to 59% compared to continuous fresh gas flow. Expiratory imposed WOB was decreased as well by 12% to 49%. In simulations of shallow to normal breathing for an adult, imposed WOB is 0.5 J l-1 at maximum. Fluctuations in mean airway pressure on account of spontaneous breathing are markedly reduced.

Conclusion: The use of the demand-flow system during HFOV results in a reduction of both imposed WOB and fluctuation in mean airway pressure. The level of imposed WOB was reduced to the physiological range of WOB. Potentially, this makes maintenance of spontaneous breathing during HFOV possible and easier in a clinical setting. Early initiation of HFOV seems more possible with this system and the possibility of weaning of patients directly on a high-frequency oscillatory ventilator is not excluded either.

Citing Articles

Management of refractory hypoxemia.

Mehta C, Mehta Y Ann Card Anaesth. 2016; 19(1):89-96.

PMID: 26750680 PMC: 4900375. DOI: 10.4103/0971-9784.173030.


CPAP and HFOV: different guises of the same underlying intensive care strategy for supporting RSV bronchiolitis.

Tasker R Intensive Care Med. 2008; 34(9):1560-1.

PMID: 18500422 DOI: 10.1007/s00134-008-1152-2.


Year in review 2006: Critical Care--Respirology.

Vasquez D, Singh J, Ferguson N Crit Care. 2007; 11(4):224.

PMID: 17764586 PMC: 2206500. DOI: 10.1186/cc5963.


Allowing for spontaneous breathing during high-frequency oscillation: the key for final success?.

Rimensberger P Crit Care. 2006; 10(4):155.

PMID: 16887013 PMC: 1750991. DOI: 10.1186/cc4993.

References
1.
French C . Work of breathing measurement in the critically ill patient. Anaesth Intensive Care. 2000; 27(6):561-73. DOI: 10.1177/0310057X9902700602. View

2.
Cereda M, Foti G, Marcora B, Gili M, Giacomini M, Sparacino M . Pressure support ventilation in patients with acute lung injury. Crit Care Med. 2000; 28(5):1269-75. DOI: 10.1097/00003246-200005000-00002. View

3.
Macintyre N, Cook D, Ely Jr E, Epstein S, FINK J, Heffner J . Evidence-based guidelines for weaning and discontinuing ventilatory support: a collective task force facilitated by the American College of Chest Physicians; the American Association for Respiratory Care; and the American College of Critical Care.... Chest. 2001; 120(6 Suppl):375S-95S. DOI: 10.1378/chest.120.6_suppl.375s. View

4.
Derdak S, Mehta S, Stewart T, Smith T, Rogers M, Buchman T . High-frequency oscillatory ventilation for acute respiratory distress syndrome in adults: a randomized, controlled trial. Am J Respir Crit Care Med. 2002; 166(6):801-8. DOI: 10.1164/rccm.2108052. View

5.
Derdak S . High-frequency oscillatory ventilation for acute respiratory distress syndrome in adult patients. Crit Care Med. 2003; 31(4 Suppl):S317-23. DOI: 10.1097/01.CCM.0000057910.50618.EB. View