Improving Mass Spectrometer Sensitivity Using a High-pressure Electrodynamic Ion Funnel Interface

Overview
Specialty Chemistry
Date 2006 Jul 15
PMID 16839773
Citations 42
Authors
Affiliations
Soon will be listed here.
Abstract

We report on a new electrodynamic ion funnel that operates at a pressure of 30 torr with no loss of ion transmission. The enhanced performance compared with previous ion funnel designs optimized for pressures of <5 torr was achieved by reducing the ion funnel capacitance and increasing the RF drive frequency (1.7 MHz) and amplitude (100-170 V peak-to-peak). No degradation of ion transmission was observed for pressures from 2 to 30 torr. The ability to operate at higher pressure enabled a new tandem ion funnel mass spectrometer interface design that can accommodate a greater gas load (e.g., from an ESI source). When combined with a multicapillary inlet, the interface provided more efficient introduction of ions, resulting in a significant enhancement in mass spectrometer sensitivity and detection limits.

Citing Articles

Implementation of Ion Mobility Spectrometry-Based Separations in Structures for Lossless Ion Manipulations (SLIM).

Hollerbach A, Conant C, Nagy G, Ibrahim Y Methods Mol Biol. 2022; 2394:453-469.

PMID: 35094340 PMC: 9526429. DOI: 10.1007/978-1-0716-1811-0_23.


Improving Signal to Noise Ratios in Ion Mobility Spectrometry and Structures for Lossless Ion Manipulations (SLIM) using a High Dynamic Range Analog-to-Digital Converter.

Hollerbach A, Giberson C, Lee J, Huntley A, Smith R, Ibrahim Y J Am Soc Mass Spectrom. 2021; 32(11):2698-2706.

PMID: 34590845 PMC: 8742676. DOI: 10.1021/jasms.1c00226.


Quantitative mass spectrometry imaging of drugs and metabolites: a multiplatform comparison.

Lamont L, Hadavi D, Viehmann B, Flinders B, Heeren R, Vreeken R Anal Bioanal Chem. 2021; 413(10):2779-2791.

PMID: 33770207 PMC: 8007509. DOI: 10.1007/s00216-021-03210-0.


A Simple Printed Circuit Board-Based Ion Funnel for Focusing Low m/z Ratio Ions with High Kinetic Energies at Elevated Pressure.

Schlottmann F, Allers M, Kirk A, Bohnhorst A, Zimmermann S J Am Soc Mass Spectrom. 2019; 30(9):1813-1823.

PMID: 31140080 DOI: 10.1007/s13361-019-02241-3.


Conformational Landscapes of Ubiquitin, Cytochrome c, and Myoglobin: Uniform Field Ion Mobility Measurements in Helium and Nitrogen Drift Gas.

May J, Jurneczko E, Stow S, Kratochvil I, Kalkhof S, McLean J Int J Mass Spectrom. 2018; 427:79-90.

PMID: 29915518 PMC: 6003721. DOI: 10.1016/j.ijms.2017.09.014.


References
1.
Smith R, Shen Y, Tang K . Ultrasensitive and quantitative analyses from combined separations-mass spectrometry for the characterization of proteomes. Acc Chem Res. 2004; 37(4):269-78. DOI: 10.1021/ar0301330. View

2.
Zhou L, Yue B, Dearden D, Lee E, Rockwood A, Lee M . Incorporation of a venturi device in electrospray ionization. Anal Chem. 2003; 75(21):5978-83. DOI: 10.1021/ac020786e. View

3.
Shaffer S, Tolmachev A, Prior D, Anderson G, Udseth H, Smith R . Characterization of an improved electrodynamic ion funnel interface for electrospray ionization mass spectrometry. Anal Chem. 1999; 71(15):2957-64. DOI: 10.1021/ac990346w. View

4.
Kim , Udseth , Smith . Improved ion transmission from atmospheric pressure to high vacuum using a multicapillary inlet and electrodynamic ion funnel interface. Anal Chem. 2000; 72(20):5014-9. DOI: 10.1021/ac0003549. View

5.
Wilm M, Mann M . Analytical properties of the nanoelectrospray ion source. Anal Chem. 1996; 68(1):1-8. DOI: 10.1021/ac9509519. View