Alterations in Regional Vascular Geometry Produced by Theoretical Stent Implantation Influence Distributions of Wall Shear Stress: Analysis of a Curved Coronary Artery Using 3D Computational Fluid Dynamics Modeling

Overview
Publisher Biomed Central
Date 2006 Jun 20
PMID 16780592
Citations 28
Authors
Affiliations
Soon will be listed here.
Abstract

Background: The success of stent implantation in the restoration of blood flow through areas of vascular narrowing is limited by restenosis. Several recent studies have suggested that the local geometric environment created by a deployed stent may influence regional blood flow characteristics and alter distributions of wall shear stress (WSS) after implantation, thereby rendering specific areas of the vessel wall more susceptible to neointimal hyperplasia and restenosis. Stents are most frequently implanted in curved vessels such as the coronary arteries, but most computational studies examining blood flow patterns through stented vessels conducted to date use linear, cylindrical geometric models. It appears highly probable that restenosis occurring after stent implantation in curved arteries also occurs as a consequence of changes in fluid dynamics that are established immediately after stent implantation.

Methods: In the current investigation, we tested the hypothesis that acute changes in stent-induced regional geometry influence distributions of WSS using 3D coronary artery CFD models implanted with stents that either conformed to or caused straightening of the primary curvature of the left anterior descending coronary artery. WSS obtained at several intervals during the cardiac cycle, time averaged WSS, and WSS gradients were calculated using conventional techniques.

Results: Implantation of a stent that causes straightening, rather than conforms to the natural curvature of the artery causes a reduction in the radius of curvature and subsequent increase in the Dean number within the stented region. This straightening leads to modest skewing of the velocity profile at the inlet and outlet of the stented region where alterations in indices of WSS are most pronounced. For example, time-averaged WSS in the proximal portion of the stent ranged from 8.91 to 11.7 dynes/cm2 along the pericardial luminal surface and 4.26 to 4.88 dynes/cm2 along the myocardial luminal surface of curved coronary arteries as compared to 8.31 dynes/cm2 observed throughout the stented region of a straight vessel implanted with an equivalent stent.

Conclusion: The current results predicting large spatial and temporal variations in WSS at specific locations in curved arterial 3D CFD simulations are consistent with clinically observed sites of restenosis. If the findings of this idealized study translate to the clinical situation, the regional geometry established immediately after stent implantation may predispose portions of the stented vessel to a higher risk of neointimal hyperplasia and subsequent restenosis.

Citing Articles

Topology optimization of coronary artery stent considering structural and hemodynamic parameters.

Ahadi F, Azadi M, Biglari M, Bodaghi M Heliyon. 2024; 10(20):e39452.

PMID: 39469694 PMC: 11513526. DOI: 10.1016/j.heliyon.2024.e39452.


Biomechanical factors and atherosclerosis localization: insights and clinical applications.

Bacigalupi E, Pizzicannella J, Rigatelli G, Scorpiglione L, Foglietta M, Rende G Front Cardiovasc Med. 2024; 11:1392702.

PMID: 39119184 PMC: 11306036. DOI: 10.3389/fcvm.2024.1392702.


Application of physics-based flow models in cardiovascular medicine: Current practices and challenges.

Vardhan M, Randles A Biophys Rev (Melville). 2024; 2(1):011302.

PMID: 38505399 PMC: 10903374. DOI: 10.1063/5.0040315.


Near-wall hemodynamic changes in subclavian artery perfusion induced by retrograde inner branched thoracic endograft implantation.

Yoon W, Mani K, Han S, Lee C, Cho J, Wanhainen A JVS Vasc Sci. 2023; 4:100116.

PMID: 37496886 PMC: 10366580. DOI: 10.1016/j.jvssci.2023.100116.


Computational Study of Hemodynamic Field of an Occluded Artery Model with Anastomosis.

Parissis P, Romeos A, Giannadakis A, Kalarakis A, Peroulis M Bioengineering (Basel). 2023; 10(2).

PMID: 36829640 PMC: 9952429. DOI: 10.3390/bioengineering10020146.


References
1.
Kute S, Vorp D . The effect of proximal artery flow on the hemodynamics at the distal anastomosis of a vascular bypass graft: computational study. J Biomech Eng. 2001; 123(3):277-83. DOI: 10.1115/1.1374203. View

2.
Wentzel J, Krams R, Schuurbiers J, Oomen J, Kloet J, van der Giessen W . Relationship between neointimal thickness and shear stress after Wallstent implantation in human coronary arteries. Circulation. 2001; 103(13):1740-5. DOI: 10.1161/01.cir.103.13.1740. View

3.
Tada S, Tarbell J . Flow through internal elastic lamina affects shear stress on smooth muscle cells (3D simulations). Am J Physiol Heart Circ Physiol. 2002; 282(2):H576-84. DOI: 10.1152/ajpheart.00751.2001. View

4.
Murata T, Hiro T, Fujii T, Yasumoto K, Murashige A, Kohno M . Impact of the cross-sectional geometry of the post-deployment coronary stent on in-stent neointimal hyperplasia: an intravascular ultrasound study. Circ J. 2002; 66(5):489-93. DOI: 10.1253/circj.66.489. View

5.
Danenberg H, Welt F, Walker 3rd M, Seifert P, Toegel G, Edelman E . Systemic inflammation induced by lipopolysaccharide increases neointimal formation after balloon and stent injury in rabbits. Circulation. 2002; 105(24):2917-22. DOI: 10.1161/01.cir.0000018168.15904.bb. View