Brain Magnetic Resonance Imaging at 3 Tesla Using BLADE Compared with Standard Rectilinear Data Sampling

Overview
Journal Invest Radiol
Specialty Radiology
Date 2006 Jun 15
PMID 16772852
Citations 16
Authors
Affiliations
Soon will be listed here.
Abstract

Objectives: We sought to evaluate Periodically Rotated Overlapping ParallEL Lines with Enhanced Reconstruction (PROPELLER; BLADE) data acquisition in comparison with standard k-space sampling techniques for axial and sagittal brain imaging at 3 T regarding imaging artifacts.

Material And Methods: Forty patients who gave consent were included in a prospective comparison of standard and PROPELLER (BLADE) k-space sampling techniques. All examinations were performed at 3 T with comparison of standard T2-weighted fluid-attenuated inversion recovery (FLAIR) to PROPELLER T2-weighted FLAIR in the axial image orientation and standard T1-weighted gradient echo to PROPELLER T1-weighted FLAIR in the sagittal image orientation. Imaging protocols were matched for spatial resolution, with data evaluation performed by 2 experienced neuroradiologists. Image data were compared regarding various image artifacts and overall image quality. Reader agreement was assessed by Cohen's kappa statistics.

Results: PROPELLER T2-weighted axial data acquisition showed significantly less pulsation and Gibb's artifacts than the standard T2-weighted scan. Even without motion correction, the frequency of ghosting (motion) artifacts was substantially lower in the PROPELLER T2-weighted data and readers concordantly (kappa = 1) rated PROPELLER as better than or equal to the standard T2-weighted scan in the majority of cases (95%; P < 0.0001). In the comparison of sagittal T1-weighted data sets, readers showed only fair agreement (kappa = 0.24) and noted consistent wrap artifacts in PROPELLER T1-weighted FLAIR.

Conclusion: PROPELLER (BLADE) brain magnetic resonance imaging is also applicable at 3 T. In addition to minimizing motion artifacts, the PROPELLER acquisition scheme reduces other magnetic resonance artifacts that would otherwise degrade scan quality.

Citing Articles

Role of high-resolution magnetic resonance imaging in preoperative tumor-node-metastasis staging evaluation of esophageal cancer: a narrative review.

Lin Y, Lin Q, Huang L, Wang J, Ma R, Liu J Quant Imaging Med Surg. 2024; 14(12):9589-9599.

PMID: 39698700 PMC: 11652015. DOI: 10.21037/qims-24-34.


Practical implications of motion correction with motion insensitive radial k-space acquisitions in MRI.

Almuqbel M, Leeper G, Palmer D, Mitchell N, Russell K, Keenan R Br J Radiol. 2018; 91(1087):20170593.

PMID: 29537305 PMC: 6221774. DOI: 10.1259/bjr.20170593.


Revised PROPELLER for T2-weighted imaging of the prostate at 3 Tesla: impact on lesion detection and PI-RADS classification.

Meier-Schroers M, Marx C, Schmeel F, Wolter K, Gieseke J, Block W Eur Radiol. 2017; 28(1):24-30.

PMID: 28687915 DOI: 10.1007/s00330-017-4949-y.


Impact of field strength and RF excitation on abdominal diffusion-weighted magnetic resonance imaging.

Riffel P, Rao R, Haneder S, Meyer M, Schoenberg S, Michaely H World J Radiol. 2013; 5(9):334-44.

PMID: 24198912 PMC: 3817292. DOI: 10.4329/wjr.v5.i9.334.


Should less motion sensitive T2-weighted BLADE TSE replace Cartesian TSE for female pelvic MRI?.

Froehlich J, Metens T, Chilla B, Hauser N, Klarhoefer M, Kubik-Huch R Insights Imaging. 2012; 3(6):611-8.

PMID: 23011875 PMC: 3505563. DOI: 10.1007/s13244-012-0193-9.