Optimizing Dialysis Delivery in Tunneled Dialysis Catheters

Overview
Journal ASAIO J
Specialty General Surgery
Date 2006 Mar 25
PMID 16557101
Citations 5
Authors
Affiliations
Soon will be listed here.
Abstract

Long-term dialysis is often delivered through tunneled central venous catheters (CVC) despite their associated morbidity and mortality rates. Because poor solute clearance might contribute to this risk, we examined the relation between blood pump speed (QB), access recirculation (AR), and dialysis delivery in patients with CVC. We conducted a prospective study on 102 patients receiving long-term hemodialysis with CVC. QB was systematically varied, and AR was measured by saline dilution with the blood supply lines in both the straight and the reversed positions during each of two dialysis sessions. During a third session, we measured ionic dialysance (EID) in patients with AR>0% and those in whom dialysis was usually delivered with catheters in the reversed position.Approximately one third (34.3%) of patients were usually run in the reversed position. Clinically significant AR (>10%) was infrequent (3%) in the straight position but common in the reversed position (86%). QB and EID were linearly correlated for both the straight and reversed positions. Approximately half (48.6%) of patients who were generally treated with their lines reversed were able to receive dialysis with the lines in the straight position within acceptable pressure limits when blood lines were switched part way through the run. In these patients, solute clearance was equivalent (EID straight 204 ml/min vs. reversed 196 ml/min, p=0.58) with lines in the straight configuration despite lower achieved QB (straight 354 ml/min vs. reversed 404 ml/min, p=0.04). Maximization of prescribed QB in CVC increased AR when blood line position was reversed but improved small solute clearance regardless of line position. This suggests that QB in CVC should be set as high as circuit pressure limits will allow, regardless of the potential for AR, and that reversal of line position confers no benefit with respect to solute clearance.

Citing Articles

Catheter Port Reversal in Citrate Continuous Veno-Venous Hemofiltration.

Boer W, van Tornout M, Vander Laenen M, Engelen K, Meex I, Jorens P Kidney Int Rep. 2021; 6(11):2775-2781.

PMID: 34805629 PMC: 8589722. DOI: 10.1016/j.ekir.2021.08.006.


Reliability of monitoring acid-base and electrolyte parameters through circuit lines during regional citrate anticoagulation-continuous renal replacement therapy.

Wang F, Dai M, Zhao Y, Yang Y, Chen Z, Lin L Nurs Crit Care. 2021; 27(5):646-651.

PMID: 34382281 PMC: 9540182. DOI: 10.1111/nicc.12696.


Computational fluid dynamics of the right atrium: Assessment of modelling criteria for the evaluation of dialysis catheters.

de Oliveira D, Owen D, Qian S, Green N, M Espino D, E T Shepherd D PLoS One. 2021; 16(2):e0247438.

PMID: 33630903 PMC: 7906423. DOI: 10.1371/journal.pone.0247438.


Reversed connection of cuffed, tunneled, dual-lumen catheters with increased blood flow rate maintains the adequacy of delivered dialysis despite the higher access recirculation.

Kousoula V, Georgianos P, Mavromatidis K, Syrganis C, Thodis E, Panagoutsos S Int Urol Nephrol. 2019; 51(10):1841-1847.

PMID: 31471844 DOI: 10.1007/s11255-019-02268-1.


Risk factors associated with hemodialysis central venous catheter malfunction; a retrospective analysis of a randomized controlled trial.

Ward D, Moist L, MacRae J, Scott-Douglas N, Zhang J, Tonelli M Can J Kidney Health Dis. 2015; 1:15.

PMID: 25960883 PMC: 4424538. DOI: 10.1186/2054-3581-1-15.