SemiSPECT: a Small-animal Single-photon Emission Computed Tomography (SPECT) Imager Based on Eight Cadmium Zinc Telluride (CZT) Detector Arrays

Overview
Journal Med Phys
Specialty Biophysics
Date 2006 Mar 15
PMID 16532954
Citations 20
Authors
Affiliations
Soon will be listed here.
Abstract

The first full single-photon emission computed tomography (SPECT) imager to exploit eight compact high-intrinsic-resolution cadmium zinc telluride (CZT) detectors, called SemiSPECT, has been completed. Each detector consists of a CZT crystal and a customized application-specific integrated circuit (ASIC). The CZT crystal is a 2.7 cm x 2.7 cm x -0.2 cm slab with a continuous top electrode and a bottom electrode patterned into a 64 x 64 pixel array by photolithography. The ASIC is attached to the bottom of the CZT crystal by indium-bump bonding. A bias voltage of -180 V is applied to the continuous electrode. The eight detectors are arranged in an octagonal lead-shielded ring. Each pinhole in the eight-pinhole aperture placed at the center of the ring is matched to each individual detector array. An object is imaged onto each detector through a pinhole, and each detector is operated independently with list-mode acquisition. The imaging subject can be rotated about a vertical axis to obtain additional angular projections. The performance of SemiSPECT was characterized using 99mTc. When a 0.5 mm diameter pinhole is used, the spatial resolution on each axis is about 1.4 mm as estimated by the Fourier crosstalk matrix, which provides an algorithm-independent average resolution over the field of view. The energy resolution achieved by summing neighboring pixel signals in a 3 x 3 window is about 10% full-width-at-half-maximum of the photopeak. The overall system sensitivity is about 0.5 x 10(-4) with the energy window of +/-10% from the photopeak. Line-phantom images are presented to visualize the spatial resolution provided by SemiSPECT, and images of bone, myocardium, and human tumor xenografts in mice demonstrate the feasibility of preclinical small-animal studies with SemiSPECT.

Citing Articles

Performance Evaluation of a Preclinical SPECT Scanner with a Collimator Designed for Medium-Sized Animals.

Matsusaka Y, Werner R, Arias-Loza P, Nose N, Sasaki T, Chen X Mol Imaging. 2022; 2022:9810097.

PMID: 35903250 PMC: 9328189. DOI: 10.1155/2022/9810097.


Preliminary evaluation of a novel energy-resolved photon-counting gamma ray detector.

Meng L, Tan J, Spartiotis K, Schulman T Nucl Instrum Methods Phys Res A. 2017; 604(3):548-554.

PMID: 28260825 PMC: 5333787. DOI: 10.1016/j.nima.2009.02.043.


Characterization of "γ-Eye": a Low-Cost Benchtop Mouse-Sized Gamma Camera for Dynamic and Static Imaging Studies.

Georgiou M, Fysikopoulos E, Mikropoulos K, Fragogeorgi E, Loudos G Mol Imaging Biol. 2016; 19(3):398-407.

PMID: 27730469 DOI: 10.1007/s11307-016-1011-4.


Molecular Imaging in the College of Optical Sciences - An Overview of Two Decades of Instrumentation Development.

Furenlid L, Barrett H, Barber H, Clarkson E, Kupinski M, Liu Z Proc SPIE Int Soc Opt Eng. 2015; 9186.

PMID: 26236069 PMC: 4520714. DOI: 10.1117/12.2064808.


Fisher Information Analysis of Depth-of-Interaction Estimation in Double-Sided Strip Detectors.

Salcin E, Barrett H, Barber H, Takeda S, Watanabe S, Takahashi T IEEE Trans Nucl Sci. 2015; 61(3):1243-1251.

PMID: 26160984 PMC: 4494127. DOI: 10.1109/TNS.2014.2317454.


References
1.
Tai Y, Chatziioannou A, Yang Y, Silverman R, Meadors K, Siegel S . MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging. Phys Med Biol. 2003; 48(11):1519-37. DOI: 10.1088/0031-9155/48/11/303. View

2.
Kastis G, Wu M, Balzer S, Wilson D, Furenlid L, Stevenson G . Tomographic Small-Animal Imaging Using a High-Resolution Semiconductor Camera. IEEE Nucl Sci Symp Conf Rec (1997). 2015; 49(1):172-175. PMC: 4643294. DOI: 10.1109/TNS.2002.998747. View

3.
Barrett , Eskin , Barber . Charge transport in arrays of semiconductor gamma-ray detectors. Phys Rev Lett. 1995; 75(1):156-159. DOI: 10.1103/PhysRevLett.75.156. View

4.
Frey E, Gilland K, Tsui B . Application of task-based measures of image quality to optimization and evaluation of three-dimensional reconstruction-based compensation methods in myocardial perfusion SPECT. IEEE Trans Med Imaging. 2003; 21(9):1040-50. DOI: 10.1109/TMI.2002.804437. View

5.
Furenlid L, Wilson D, Chen Y, Kim H, Pietraski P, Crawford M . FastSPECT II: A Second-Generation High-Resolution Dynamic SPECT Imager. IEEE Trans Nucl Sci. 2010; 51(3):631-635. PMC: 2945369. DOI: 10.1109/TNS.2004.830975. View