Validation of the Use of Foreign Gas Rebreathing Method for Non-invasive Determination of Cardiac Output in Heart Disease Patients

Overview
Date 2005 Dec 17
PMID 16358372
Citations 14
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: To compare a new device (Innocor) for non-invasive measurement of cardiac output (CO) by foreign gas rebreathing method with conventional techniques used in the measurements of cardiac function.

Methods: Cardiac outputs measured by Innocor (CO(RB)) were compared with CO obtained by echocardiography (CO(EC)), Swan-Ganz thermodilution (CO(TD)), and left ventricle radiography (CO(LVR)) in 34 patients subjected to cardiac catheterization. Values obtained from the four methods were analyzed by linear regression and paired values were compared by the method of Bland and Altman in SPSS.

Results: There was strong positive correlation (r=0.94) between Innocor cardiac output values and the corresponding values obtained by thermodilution and between CO(EC) and CO(LVR) values. Thermodilution appears to overestimate cardiac output when compared to the values obtained with Innocor by (0.66+/-0.22) L/min (P<0.0001). There was no correlation between data obtained by Innocor and the corresponding CO(EC) and CO(LVR) values.

Conclusion: Innocor CO(RB) is an easy, safe and well established method for non-invasive measurement of cardiac output with good prospects for clinical application in heart disease patients.

Citing Articles

Mechanism involved of post-exercise cold water immersion: Blood redistribution and increase in energy expenditure during rewarming.

Giraud D, Pomportes L, Nicol C, Bertin D, Gardarein J, Hays A Temperature (Austin). 2024; 11(2):137-156.

PMID: 38846524 PMC: 11152100. DOI: 10.1080/23328940.2024.2303332.


Non-Invasive Cardiac Output Measurement Using Inert Gas Rebreathing Method during Cardiopulmonary Exercise Testing-A Systematic Review.

Chwiedz A, Minarowski L, Mroz R, Hady H J Clin Med. 2023; 12(22).

PMID: 38002766 PMC: 10671909. DOI: 10.3390/jcm12227154.


Non-invasive measurement of hemodynamic response to postural stress using inert gas rebreathing.

Stach K, Michels J, Doesch C, Brade J, Papavassiliu T, Borggrefe M Biomed Rep. 2019; 11(3):98-102.

PMID: 31423303 PMC: 6684938. DOI: 10.3892/br.2019.1229.


Validation of intracardiac shunt using thoracic bioimpedance and inert gas rebreathing in adults before and after percutaneous closure of atrial septal defect in a cardiology research unit: study protocol.

Filaire L, Chalard A, Perrault H, Tresorier R, Lusson J, Pereira B BMJ Open. 2019; 9(5):e024389.

PMID: 31133575 PMC: 6538205. DOI: 10.1136/bmjopen-2018-024389.


Evaluation of inert gas rebreathing for determination of cardiac output: influence of age, gender and body size.

Middlemiss J, Cocks A, Paapstel K, Maki-Petaja K, Sunita , Wilkinson I Hypertens Res. 2018; 42(6):834-844.

PMID: 30560890 PMC: 8076049. DOI: 10.1038/s41440-018-0179-1.


References
1.
Gabrielsen A, Videbaek R, Schou M, Damgaard M, Kastrup J, Norsk P . Non-invasive measurement of cardiac output in heart failure patients using a new foreign gas rebreathing technique. Clin Sci (Lond). 2002; 102(2):247-52. View

2.
SACKNER M, Greeneltch D, Heiman M, Epstein S, Atkins N . Diffusing capacity, membrane diffusing capacity, capillary blood volume, pulmonary tissue volume, and cardiac output measured by a rebreathing technique. Am Rev Respir Dis. 1975; 111(2):157-65. DOI: 10.1164/arrd.1975.111.2.157. View

3.
Sweet R, Moraski R, Russell Jr R, Rackley C . Relationship between echocardiography, cardiac output, and abnormally contracting segments in patients with ischemic heart disease. Circulation. 1975; 52(4):634-41. DOI: 10.1161/01.cir.52.4.634. View

4.
Petrini M, Peterson B, HYDE R . Lung tissue volume and blood flow by rebreathing theory. J Appl Physiol Respir Environ Exerc Physiol. 1978; 44(5):795-802. DOI: 10.1152/jappl.1978.44.5.795. View

5.
Damgaard M, Norsk P . Effects of ventilation on cardiac output determined by inert gas rebreathing. Clin Physiol Funct Imaging. 2005; 25(3):142-7. DOI: 10.1111/j.1475-097X.2005.00602.x. View