» Articles » PMID: 16237142

Brain Tumors: Full- and Half-dose Contrast-enhanced MR Imaging at 3.0 T Compared with 1.5 T--Initial Experience

Overview
Journal Radiology
Specialty Radiology
Date 2005 Oct 21
PMID 16237142
Citations 28
Authors
Affiliations
Soon will be listed here.
Abstract

Purpose: To prospectively and intraindividually compare the effect of magnetic resonance (MR) imaging at a higher magnetic field strength (3.0 T) on contrast-to-noise ratio (CNR) at different doses of a T1-shortening contrast agent in patients with contrast-enhancing brain lesions, with 1.5-T MR imaging as a reference standard.

Materials And Methods: Institutional review board approval and informed consent were obtained for all patient and volunteer studies. Twelve patients (six women, six men; mean age, 58 years; range, 29-76 years) with 12 enhancing brain lesions (11 patients with primary brain tumors and one with a solitary cerebral metastasis) underwent contrast material-enhanced MR imaging three times, on three separate days: once at 1.5 T with a full dose of 0.10 mmol/kg gadopentetate dimeglumine, once at 3.0 T with a full dose, and once at 3.0 T with half that dose, 0.05 mmol/kg. The same contrast-enhanced T1-weighted spin-echo images (repetition time msec/echo time msec, 500/12; section thickness, 5 mm; matrix, 256 x 205) were obtained at both 3.0 T and 1.5 T after prior optimization of parameters at 3.0 T. The number and conspicuity of enhancing brain lesions were assessed with blinded clinical image reading. Signal-to-noise ratio and CNR were determined with region of interest analysis of enhancing lesions and normal contralateral white matter. For 3.0 T with half the standard dose and with the full dose, CNR of lesions was intraindividually compared with CNR at 1.5 T with the full dose by using the Wilcoxon matched-pairs signed rank test.

Results: At 3.0 T and full dose, CNR was 2.8-fold higher than that at 1.5 T and full dose (P < .001). At the same time, higher lesion conspicuity at clinical image reading was observed. With only half the standard dose, MR imaging at 3.0 T still yielded higher CNR (1.3-fold higher) than that with full dose at 1.5 T (P < .01).

Conclusion: With the same amount of contrast agent, MR imaging at 3.0 T offered a significantly higher CNR of enhancing cerebral lesions, compared with that at 1.5 T; even with the dose reduced by half, CNR was still higher at 3.0 T.

Citing Articles

Improved detection of brain metastases using contrast-enhanced 3D black-blood TSE sequences compared to post-contrast 3D T1 GRE: a comparative study on 1.5-T MRI.

Chen Zhou Z, Salvador Alvarez E, Hilario A, Cardenas Del Carre A, Romero Coronado J, Lechuga C Eur Radiol. 2025; .

PMID: 39841203 DOI: 10.1007/s00330-025-11363-0.


The feasibility of half-dose contrast-enhanced scanning of brain tumours at 5.0 T: a preliminary study.

Jiang Z, Sun W, Xu D, Mei H, Yuan J, Song X BMC Med Imaging. 2024; 24(1):88.

PMID: 38615005 PMC: 11016225. DOI: 10.1186/s12880-024-01270-z.


Review of strategies to reduce the contamination of the water environment by gadolinium-based contrast agents.

Dekker H, Stroomberg G, van der Molen A, Prokop M Insights Imaging. 2024; 15(1):62.

PMID: 38411847 PMC: 10899148. DOI: 10.1186/s13244-024-01626-7.


Brain Metastases From Lung Adenocarcinoma May Preferentially Involve the Distal Middle Cerebral Artery Territory and Cerebellum.

Kwon H, Kim J, Park M, Kim J, Kim M, Suh S Front Oncol. 2020; 10:1664.

PMID: 32984041 PMC: 7484698. DOI: 10.3389/fonc.2020.01664.


Brain Tumor-Enhancement Visualization and Morphometric Assessment: A Comparison of MPRAGE, SPACE, and VIBE MRI Techniques.

Danieli L, Riccitelli G, Distefano D, Prodi E, Ventura E, Cianfoni A AJNR Am J Neuroradiol. 2019; 40(7):1140-1148.

PMID: 31221635 PMC: 7048542. DOI: 10.3174/ajnr.A6096.