Pulmonary Function Electronic Monitoring Devices: a Randomized Agreement Study
Overview
Authors
Affiliations
Study Objectives: To compare in a clinical setting the within-session reproducibility of two pulmonary function electronic monitoring devices (PiKo-1; Ferraris Respiratory Europe; Hereford, UK; and Spirotel; MIR; Rome, Italy) with one mechanical device (Mini-Wright Peak-Flow Meter; Clement-Clarke International; Harlow, Essex, UK), and to evaluate the accuracy of these devices using as reference an office pneumotachograph.
Design, Setting, And Participants: After detailed instructions, adults without airways diseases and patients with stable asthma attending an outpatient clinic performed four sets of expiratory maneuvers, one set for each device, in a strictly random order. Each set comprised three maneuvers with 2 to 3 min of rest between them.
Measurements: Reproducibility of FEV1 and peak expiratory flow (PEF) was assessed by a coefficient of variation (CV) and intraclass correlation coefficient (ICC), and accuracy was assessed by ICC and limits of agreement.
Results: Of the 38 participants evaluated, 71% were women and 61% had asthma. Ages ranged from 18 to 58 years, and FEV1 ranged from 1.2 to 4.8 L. In all monitoring devices, CV was < 6% and ICC was > 0.94 for the reproducibility of both FEV1 and PEF measurements. The accuracy of the PiKo-1 device was better for FEV1 (ICC = 0.98) than for PEF (ICC = 0.90). The Spirotel device had similar results for FEV1 and PEF (ICC = 0.95). The Mini-Wright device had the lowest accuracy (ICC = 0.87), particularly for PEF values < 500 L/min.
Conclusions: These low-cost and easy-to-use electronic monitoring devices showed a very good reproducibility and were in agreement with the pneumotachograph. Therefore, the PiKo-1 and Spirotel devices seem adequate for both screening and monitoring. However, prospective studies are still needed to assess their long-term reproducibility and usability and, particularly, the effects on the improvement of respiratory care.
Khor Y, Poberezhets V, Buhr R, Chalmers J, Choi H, Fan V Am J Respir Crit Care Med. 2024; 211(2):174-193.
PMID: 39585746 PMC: 11812536. DOI: 10.1164/rccm.202410-2080ST.
Poon E, Sun F, Tse A, Tsang J, Chung A, Lai Y J Exerc Sci Fit. 2024; 22(2):134-139.
PMID: 38347889 PMC: 10859274. DOI: 10.1016/j.jesf.2024.01.003.
Junge M, Kruger M, Wahner-Roedler D, Bauer B, Dorr M, Bahls M BMC Med Res Methodol. 2023; 23(1):103.
PMID: 37095457 PMC: 10127382. DOI: 10.1186/s12874-023-01911-x.
Fossat G, Desmalles E, Courtes L, Fossat C, Boulain T Respir Care. 2023; 68(4):470-477.
PMID: 36878644 PMC: 10173113. DOI: 10.4187/respcare.10412.
NCS Assessments of the Motor, Sensory, and Physical Health Domains.
Clark J, Pate R, Rine R, Christy J, Dalton P, Damiano D Front Pediatr. 2021; 9:622542.
PMID: 34900852 PMC: 8661476. DOI: 10.3389/fped.2021.622542.