Patellofemoral Forces After Total Knee Arthroplasty: Effect of Extensor Moment Arm

Overview
Journal Knee
Publisher Elsevier
Specialty Orthopedics
Date 2005 Mar 8
PMID 15749440
Citations 28
Authors
Affiliations
Soon will be listed here.
Abstract

Total knee arthroplasty implant designs with larger extensor moment arms theoretically should generate lower extensor forces for the same externally applied loads. This study measured knee kinematics, quadriceps forces, and patellofemoral forces under conditions of dynamic knee extension under load in two knee designs with differing quadriceps moment arms. Six human cadaver knees were tested both before implantation and after sequential implantation with two posterior cruciate retaining designs. The extensor moment arm of the LMA (long extensor moment arm design, Scorpio, Howmedica Osteonics, Rutherford, NJ) was approximately 1 cm longer than that of the Control design (7000, Howmedica Osteonics). Quadriceps tension was measured during dynamic closed kinetic chain knee extension. Patellar compressive and shear forces were also recorded using a patellar component instrumented with a custom triaxial load transducer. Knee kinematics were monitored using a three-dimensional electromagnetic tracking device. Both designs produced similar patterns of femoral rollback and tibial rotation. Quadriceps tension was lower in the LMA design compared with the Control design. Patellofemoral compressive forces were also significantly reduced in the LMA design when compared with Control (8-18% lower at angles greater than 50 degrees flexion). The design with the longer extensor moment arm required less quadriceps force to extend the knee under load and reduced patellofemoral compressive forces. Reduced quadriceps forces may facilitate postoperative rehabilitation and activities such as stair climbing. Reduction in patellofemoral forces could reduce patellar complications such as anterior knee pain, component wear, and loosening.

Citing Articles

Impact of the external knee flexion moment on patello-femoral loading derived from loads and kinematics.

Trepczynski A, Kneifel P, Heyland M, Leskovar M, Moewis P, Damm P Front Bioeng Biotechnol. 2025; 12:1473951.

PMID: 39881960 PMC: 11774905. DOI: 10.3389/fbioe.2024.1473951.


Evaluation of the Tibial Plateau-Patella Angle (TPPA) in Dogs.

Zaimovic N, Lorinson D, Lorinson K, Tichy A, Bockstahler B Animals (Basel). 2024; 14(12).

PMID: 38929417 PMC: 11200802. DOI: 10.3390/ani14121798.


A Rare Presentation of Patella Button Aseptic Loosening After a Total Knee Replacement Without Evidence of Radiographic Loosening.

Teh H, Kassim A, Chopra S, Selvaratnam V Cureus. 2023; 15(2):e34665.

PMID: 36909109 PMC: 9997687. DOI: 10.7759/cureus.34665.


Correlation between component alignment and short-term clinical outcomes after total knee arthroplasty.

Luan Y, Zhang M, Ran T, Wang H, Fang C, Nie M Front Surg. 2022; 9:991476.

PMID: 36311927 PMC: 9606652. DOI: 10.3389/fsurg.2022.991476.


Comparison of Single-Radius with Multiple-Radius Femur in Total Knee Arthroplasty: A Meta-Analysis of Prospective Randomized Controlled Trials.

Lei T, Jiang Z, Qian H, Backstein D, Lei P, Hu Y Orthop Surg. 2022; 14(9):2085-2095.

PMID: 35924690 PMC: 9483041. DOI: 10.1111/os.13391.