Mechanical Behavior of Screws in Normal and Osteoporotic Bone

Overview
Journal Osteoporos Int
Date 2004 Nov 2
PMID 15517184
Citations 28
Authors
Affiliations
Soon will be listed here.
Abstract

Fracture fixation in severe osteoporotic bone by means of implants that rely on screw anchorage is still a clinical problem. So far, a sufficiently accurate prediction of the holding capacity of screws as a function of local bone morphology has not been obtained. In this study the ultimate pullout loads of screws in the epi-, meta-, and diaphyseal regions of human tibiae were correlated to the cortical thicknesses and cancellous bone mineral densities at the screw axes determined from QCT densitometric data. Stepwise multiple linear regression showed that in regions with cortical thicknesses below 1.5 mm, cancellous density determined the ultimate pullout load (R2 = 0.85, p < 0.001), while in regions with cortices above 1.5 mm, cortical thickness alone significantly influenced the holding capacity of a screw (R2 = 0.90, p < 0.001). The findings of this study provide a basis for a bone morphology-related pre-operative estimation of the holding capacity of screws, which could help to improve their proper application in osteoporotic bone.

Citing Articles

Zero-profile implant system versus novel plate systems after ACDF for comparison of sagittal balance parameters and clinical efficacy analysis.

Gong Y, Zhuo H, Zhou Z, Cheng Z, Gan Y, He J J Orthop Surg Res. 2024; 19(1):363.

PMID: 38898467 PMC: 11186272. DOI: 10.1186/s13018-024-04857-y.


Impact of osteoporosis and osteoporosis medications on fracture healing: a narrative review.

Chandran M, Akesson K, Javaid M, Harvey N, Blank R, Brandi M Osteoporos Int. 2024; 35(8):1337-1358.

PMID: 38587674 PMC: 11282157. DOI: 10.1007/s00198-024-07059-8.


Comparing the biomechanical properties of conventional suture and all-suture anchors using patient-specific and realistic osteoporotic and non-osteoporotic phantom using 3D printing.

Ock J, Seo J, Koh K, Kim N Sci Rep. 2023; 13(1):20976.

PMID: 38017064 PMC: 10684536. DOI: 10.1038/s41598-023-48392-y.


On the material dependency of peri-implant morphology and stability in healing bone.

Bruns S, Kruger D, Galli S, Wieland D, Hammel J, Beckmann F Bioact Mater. 2023; 28:155-166.

PMID: 37250865 PMC: 10212791. DOI: 10.1016/j.bioactmat.2023.05.006.


The ideal site of cement application in cement augmented sacroiliac screw fixation: the biomechanical perspective.

Albers C, Zderic I, Kastner P, Gueorguiev B, Tosounidis T, Keel M Eur J Trauma Emerg Surg. 2022; 49(3):1449-1458.

PMID: 36508022 PMC: 10229737. DOI: 10.1007/s00068-022-02187-4.


References
1.
Faran K, Ichioka N, Trzeciak M, Han S, Medige J, Moy O . Effect of bone quality on the forces generated by compression screws. J Biomech. 1999; 32(8):861-4. DOI: 10.1016/s0021-9290(99)00076-7. View

2.
Snyder S, Schneider E . Estimation of mechanical properties of cortical bone by computed tomography. J Orthop Res. 1991; 9(3):422-31. DOI: 10.1002/jor.1100090315. View

3.
Messmer P, Long G, Suhm N, Regazzoni P, Jacob A . Volumetric model determination of the tibia based on 2D radiographs using a 2D/3D database. Comput Aided Surg. 2002; 6(4):183-94. DOI: 10.1002/igs.10009. View

4.
Harnroongroj T, Techataweewan A . Determination of the role of the cancellous bone in generation of screw holding power at metaphysis. Clin Biomech (Bristol). 1999; 14(5):364-6. DOI: 10.1016/s0268-0033(98)00107-7. View

5.
Pearson J, Ruegsegger P, Dequeker J, Henley M, Bright J, Reeve J . European semi-anthropomorphic phantom for the cross-calibration of peripheral bone densitometers: assessment of precision accuracy and stability. Bone Miner. 1994; 27(2):109-20. DOI: 10.1016/s0169-6009(08)80213-9. View