Stroke Volume and Systolic Time Intervals: Beat-to-beat Comparison Between Echocardiography and Ambulatory Impedance Cardiography in Supine and Tilted Positions

Overview
Publisher Springer
Date 2004 Oct 27
PMID 15503973
Citations 8
Authors
Affiliations
Soon will be listed here.
Abstract

The aim of the study was to compare stroke volume (SV), ejection time (ET) and pre-ejection period (PEP) measurements obtained using a central haemodynamics ambulatory monitoring device based on impedance cardiography (ICG), in supine and tilted positions (60 degrees), with pulsed Doppler echocardiography as a non-invasive reference method. The Holter-type ICG device was used for off-line, beat-to-beat, automatic determination of SV, ET and PEP. ICG data were compared with those obtained simultaneously using pulsed Doppler echocardiography in the ascending aorta from a suprasternal projection, 1 min before and 10 min after tilting. The tests were performed in 13 young, healthy subjects (six men and seven women, aged 23-33 years). Linear regression between the measured values obtained for all subjects was described by the following formulas: SVicg= 13.9 + 0.813 x SVecho (r = 0.857, SEE = 9.03, n = 496), ETicg = 16.8 + 0.987 x ETecho (r = 0.841, SEE=21.3, n = 496), PEPicg= 22.8 + 0.890 x PEPecho (r = 0.727, SEE = 14.6, n = 496). The data showed that ambulatory impedance cardiography gives useful absolute values of SV and systolic time intervals measured in supine and tilted positions.

Citing Articles

A Review on Recent Advancements of Biomedical Radar for Clinical Applications.

Dong S, Wen L, Ye Y, Zhang Z, Wang Y, Liu Z IEEE Open J Eng Med Biol. 2024; 5:707-724.

PMID: 39184961 PMC: 11342929. DOI: 10.1109/OJEMB.2024.3401105.


Perioperative advanced haemodynamic monitoring of patients undergoing multivisceral debulking surgery: an observational pilot study.

Middel C, Stetzuhn M, Sander N, Kalkbrenner B, Tigges T, Pielmus A Intensive Care Med Exp. 2023; 11(1):61.

PMID: 37682496 PMC: 10491568. DOI: 10.1186/s40635-023-00543-1.


Non-Invasive Wearable Patch Utilizing Seismocardiography for Peri-Operative Use in Surgical Patients.

Semiz B, Carek A, Johnson J, Ahmad S, Heller J, Vicente F IEEE J Biomed Health Inform. 2020; 25(5):1572-1582.

PMID: 33090962 PMC: 8189504. DOI: 10.1109/JBHI.2020.3032938.


Blood Pressure Estimation Using On-body Continuous Wave Radar and Photoplethysmogram in Various Posture and Exercise Conditions.

Pour Ebrahim M, Heydari F, Wu T, Walker K, Joe K, Redoute J Sci Rep. 2019; 9(1):16346.

PMID: 31705001 PMC: 6841972. DOI: 10.1038/s41598-019-52710-8.


Automatic Detection of Seismocardiogram Sensor Misplacement for Robust Pre-Ejection Period Estimation in Unsupervised Settings.

Ashouri H, Inan O IEEE Sens J. 2017; 17(12):3805-3813.

PMID: 29085256 PMC: 5659316. DOI: 10.1109/JSEN.2017.2701349.


References
1.
Francis D, Coats A, Gibson D . How high can a correlation coefficient be? Effects of limited reproducibility of common cardiological measures. Int J Cardiol. 1999; 69(2):185-9. DOI: 10.1016/s0167-5273(99)00028-5. View

2.
Milsom I, SIVERTSSON R, Biber B, Olsson T . Measurement of stroke volume with impedance cardiography. Clin Physiol. 1982; 2(5):409-17. DOI: 10.1111/j.1475-097x.1982.tb00047.x. View

3.
Charloux A, Lonsdorfer-Wolf E, Richard R, Lampert E, Oswald-Mammosser M, Mettauer B . A new impedance cardiograph device for the non-invasive evaluation of cardiac output at rest and during exercise: comparison with the "direct" Fick method. Eur J Appl Physiol. 2000; 82(4):313-20. DOI: 10.1007/s004210000226. View

4.
Smorawinski J, Nazar K, Kaciuba-Uscilko H, Kaminska E, Cybulski G, Kodrzycka A . Effects of 3-day bed rest on physiological responses to graded exercise in athletes and sedentary men. J Appl Physiol (1985). 2001; 91(1):249-57. DOI: 10.1152/jappl.2001.91.1.249. View

5.
Zoghbi W, Quinones M . Determination of cardiac output by Doppler echocardiography: a critical appraisal. Herz. 1986; 11(5):258-68. View