» Articles » PMID: 15100857

An Integrated Microfluidic Biochemical Detection System for Protein Analysis with Magnetic Bead-based Sampling Capabilities

Overview
Journal Lab Chip
Specialties Biotechnology
Chemistry
Date 2004 Apr 22
PMID 15100857
Citations 58
Authors
Affiliations
Soon will be listed here.
Abstract

This paper presents the development and characterization of an integrated microfluidic biochemical detection system for fast and low-volume immunoassays using magnetic beads, which are used as both immobilization surfaces and bio-molecule carriers. Microfluidic components have been developed and integrated to construct a microfluidic biochemical detection system. Magnetic bead-based immunoassay, as a typical example of biochemical detection and analysis, has been successfully performed on the integrated microfluidic biochemical analysis system that includes a surface-mounted biofilter and electrochemical sensor on a glass microfluidic motherboard. Total time required for an immunoassay was less than 20 min including sample incubation time, and sample volume wasted was less than 50 microl during five repeated assays. Fast and low-volume biochemical analysis has been successfully achieved with the developed biofilter and immunosensor, which is integrated to the microfluidic system. Such a magnetic bead-based biochemical detection system, described in this paper, can be applied to protein analysis systems.

Citing Articles

APHRODITE: A Compact Lab-on-Chip Biosensor for the Real-Time Analysis of Salivary Biomarkers in Space Missions.

Nardi L, Davis N, Sansolini S, Baratto de Albuquerque T, Laarraj M, Caputo D Biosensors (Basel). 2024; 14(2).

PMID: 38391991 PMC: 10887022. DOI: 10.3390/bios14020072.


Shaking Device for Homogeneous Dispersion of Magnetic Beads in Droplet Microfluidics.

Poles M, Meggiolaro A, Cremaschini S, Marinello F, Filippi D, Pierno M Sensors (Basel). 2023; 23(12).

PMID: 37420565 PMC: 10304097. DOI: 10.3390/s23125399.


Attogram-level light-induced antigen-antibody binding confined in microflow.

Iida T, Hamatani S, Takagi Y, Fujiwara K, Tamura M, Tokonami S Commun Biol. 2022; 5(1):1053.

PMID: 36203087 PMC: 9537419. DOI: 10.1038/s42003-022-03946-0.


Detection of anti-p53 autoantibodies in saliva using microfluidic chips for the rapid screening of oral cancer.

Lin Y, Wu C, Peng Y, Wu C, Chang Y, Chang K RSC Adv. 2022; 8(28):15513-15521.

PMID: 35539469 PMC: 9080182. DOI: 10.1039/c7ra13734f.


High-throughput precise particle transport at single-particle resolution in a three-dimensional magnetic field for highly sensitive bio-detection.

Abedini-Nassab R, Shourabi R Sci Rep. 2022; 12(1):6380.

PMID: 35430583 PMC: 9013386. DOI: 10.1038/s41598-022-10122-1.