Strategies for Formulating Appropriate MDCT Techniques when Imaging the Chest, Abdomen, and Pelvis in Pediatric Patients

Overview
Specialties Oncology
Radiology
Date 2004 Mar 25
PMID 15039151
Citations 25
Authors
Affiliations
Soon will be listed here.
Abstract

Objective: Our aim was to formulate appropriate MDCT chest and abdominopelvic CT scan protocols for pediatric patients.

Materials And Methods: Surface radiation dose measurements from a set of anthropomorphic phantoms (nominal 1 year old, 5 year old, and 10 year old) and an adult phantom were compared with standard CT dose index measurements. Image-noise values on axial 5-mm-thick anthropomorphic phantom images were obtained as a measure of image quality.

Results: Peripheral CT dose index values obtained with the standard 16-cm acrylic phantom were within approximately 10% of the CT surface dose measurements for the pediatric anthropomorphic phantoms for both chest and abdominopelvic scan protocols. The noise value for the adult phantom image acquired using a typical clinical CT technique was identified, and targeting this level of noise for pediatric CT examinations resulted in a decrease in dose of 60-90%. Initially, 80 kVp was selected for use with very small children; however, beam-hardening artifacts were severe enough to cause us to abandon this option. Current pediatric protocols at M. D. Anderson Cancer Center rely on 100- and 120-kVp settings. The display field-of-view parameter can be used as a surrogate for patient size to develop clinical pediatric CT protocol charts.

Conclusion: CT dose index measurements obtained using the 16-cm standard acrylic phantom are sufficiently accurate for estimating chest and abdominopelvic CT entrance exposures for pediatric patients of the same approximate size as the anthropomorphic phantoms used in this study. Image-noise measurements can be used to adjust chest and abdominopelvic CT techniques for pediatric populations, resulting in a decrease in measured entrance dose by 60-90%.

Citing Articles

A Study of CT-derived Radiation Dose Calculation in Lung Q-SPECT/CT Imaging.

Silov G, Boga F Mol Imaging Radionucl Ther. 2023; 32(3):214-220.

PMID: 37870282 PMC: 10600546. DOI: 10.4274/mirt.galenos.2023.55707.


Validation of computed tomography angiography as a complementary test in the assessment of renal artery stenosis: a comparison with digital subtraction angiography.

Fleury A, Durand R, Cahill A, Zhu X, Meyers K, Otero H Pediatr Radiol. 2021; 51(13):2507-2520.

PMID: 34374838 DOI: 10.1007/s00247-021-05145-1.


Evaluation of image quality and radiation dose in low tube voltage coronary computed tomography angiography.

Mousavi-Gazafroudi S, Sajjadieh-Khajouei A, Moradi M, Mousavi-Gazafroudi S, Yadegarfar G, Tavakoli M ARYA Atheroscler. 2020; 15(5):205-210.

PMID: 31949446 PMC: 6954360. DOI: 10.22122/arya.v15i5.1406.


Application of low-tube current with iterative model reconstruction on Philips Brilliance iCT Elite FHD in the accuracy of spinal QCT using a European spine phantom.

Wu Y, Jiang Y, Han X, Wang M, Gao J Quant Imaging Med Surg. 2018; 8(1):32-38.

PMID: 29541621 PMC: 5835662. DOI: 10.21037/qims.2018.02.03.


Size-based quality-informed framework for quantitative optimization of pediatric CT.

Samei E, Li X, Frush D J Med Imaging (Bellingham). 2017; 4(3):031209.

PMID: 28840168 PMC: 5565677. DOI: 10.1117/1.JMI.4.3.031209.